MELODY TRANSCATHETER PULMONARY VALVE
Report
- Report Number
- 2025587-2021-00603
- Event Type
- Injury
- Date Received
- February 18, 2021
- Date of Event
- January 23, 2021
- Report Date
- February 22, 2021
- Manufacturer
- MEDTRONIC HEART VALVES DIVISION
- Product Code
- NPV
- PMA / PMN Number
- P140017
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
PATIENT WEIGHT AND DEVICE SERIAL NUMBER WERE PROVIDED AND HAVE BEEN ADDED TO THIS REPORT. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
CITATION: KOGURE T, BUTERA G, MICHIELON G, KABIR S, REGAN W, JONES MI, ROSENTHAL E, QURESHI SA. MELODY VALVE FRACTURE CAUSING MITRAL STENOSIS: NOVEL SOLUTION FOR TRANSAPICAL VALVE IN VALVE. ANN THORAC SURG. 2021 JAN 23:S0003-4975(21)00115-6. DOI: 10.1016/J.ATHORACSUR.2020.12.051. EARLIEST DATE OF PUBLISH USED FOR DATE OF EVENT. NO UNIQUE DEVICE IDENTIFIER (SERIAL/LOT) NUMBERS WERE PROVIDED; WITHOUT THIS INFORMATION IT COULD NOT BE DETERMINED WHETHER THESE OBSERVATIONS HAVE BEEN PREVIOUSLY REPORTED. WITHOUT RETURN OF THE PRODUCT NO DEFINITIVE CONCLUSION CAN BE MADE REGARDING THE CLINICAL OBSERVATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION VIA LITERATURE REGARDING A (B)(6)-YEAR-OLD MALE PATIENT WITH A TRANSCATHETER BIOPROSTHETIC PULMONARY VALVE IMPLANTED IN THE MITRAL POSITION. FOUR YEARS AFTER THE IMPLANT DYSPNEA AND PULMONARY EDEMA WERE NOTED. A TRANSTHORACIC ECHOCARDIOGRAM (TTE) INDICATED SEVERE MITRAL STENOSIS, INCREASED GRADIENT MEASUREMENTS (22MMHG). A TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) INDICATED MULTIPLE STENT FRACTURES WITH A MOBILE SEGMENT IN THE INFLOW PORTION OF THE VALVE. SUBSEQUENTLY A SUCCESSFUL VALVE-IN-VALVE WAS PERFORMED VIA A MINI-THORACOTOMY AND LEFT VENTRICULAR APICAL ACCESS. NO ADDITIONAL ADVERSE PATIENT EFFECTS OR PRODUCT PERFORMANCE ISSUES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243132 | MELODY TRANSCATHETER PULMONARY VALVE | PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED | NPV | MEDTRONIC HEART VALVES DIVISION | PB1018 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR | Required Intervention |