FDA Adverse Event
Malfunction
Summary report: N
PROTEUS XR/A
MDR report key: 1133730
·
Received August 20, 2008
Report
- Report Number
- 9613445-2008-00022
- Event Type
- Malfunction
- Date Received
- August 20, 2008
- Date of Event
- June 12, 2008
- Report Date
- June 12, 2008
- Manufacturer
- GE HANGWEI MEDICAL SYSTEMS CO., LTD.
- Product Code
- KPR
- PMA / PMN Number
- K993090
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE GE FIELD ENGINEER (FE) EVALUATED THE SYSTEM AND FOUND MISADJUSTED OPTICAL SENSOR FLAG THAT DEACTIVATED THE LOCKS. THE FE ADJUSTED THE SENSOR FLAG AND VERIFIED THAT THE TABLE LOCKS WERE PERFORMING ACCORDING TO SPECIFICATIONS.
Description of Event or Problem · 1
THE TABLE LOCKS DID NOT ACTUATE, CAUSING THE TABLETOP TO MOVE IN TWO DIRECTIONS WITHOUT RESISTANCE. THERE WAS NO INJURY REPORTED. THE ISSUE WAS DISCOVERED AS THE OPERATOR WAS POSITIONING THE TABLE TOP. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING A PT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROTEUS XR/A | KPR | GE HANGWEI MEDICAL SYSTEMS CO., LTD. | 2259988 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |