FDA Adverse Event
Malfunction
Summary report: N
GE OEC 7600
MDR report key: 1133694
·
Received August 20, 2008
Report
- Report Number
- 9680959-2008-00142
- Event Type
- Malfunction
- Date Received
- August 20, 2008
- Date of Event
- August 1, 2008
- Report Date
- August 18, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE IS CURRENTLY UNDER INVESTIGATION. A FOLLOW UP REPORT WILL BE FILED WHEN ADDITIONAL DETAILS BECOME AVAILABLE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 7600 SYSTEM HAD POOR IMAGE QUALITY DURING A CASE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 7600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |