FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1133672 · Received August 20, 2008

Report

Report Number
1720753-2008-24849
Event Type
Malfunction
Date Received
August 20, 2008
Date of Event
July 31, 2008
Report Date
August 20, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE LEMO PIN CONNECTOR WAS CLEANED AND REPAIRED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM HAD A COMMUNICATION ERROR. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1