M530 OHX
Report
- Report Number
- 3003974370-2021-00003
- Event Type
- Malfunction
- Date Received
- February 17, 2021
- Date of Event
- January 19, 2021
- Report Date
- May 7, 2021
- Manufacturer
- LEICA MICROSYSTEMS (SCHWEIZ) AG
- Product Code
- EPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THIS IS A FINAL REPORT. AN INVESTIGATION OF THE REPORTED INCIDENT WHERE THE ILLUMINATION OF AN M530 OHX SHUT OFF DURING SURGERY WAS CONDUCTED. AN INITIAL INVESTIGATION PERFORMED BY THE FSE ON THE ACTUAL DEVICE REVEALED, THAT THE ILLUMINATION SHUT DOWN WITH A LAMP DEFECTIVE ERROR. THE UNIT CONTINUED TO WORK DESPITE A DEFECTIVE COMPONENT (OVERHEATED NTC "R15") ON THE BOARD. REPLACING THE VOLTAGE SELECTOR BOARD CORRECTED THE ISSUE AND THE DEVICE WAS FUNCTIONING ACCORDING TO SPECIFICATIONS. A REVIEW OF THE COMPLAINT DATABASE AND THE SPARE PART CONSUMPTION WAS PERFORMED. THE REVIEW DID NOT REVEAL INDICATIONS FOR AN ADVERSE TREND. TESTS ON A REPRESENTATIVE M530 OHX, A DESIGN REVIEW OF THE ELECTRONIC CIRCUITS INCLUDING COMPONENT SPECIFICATIONS AND AN INVESTIGATION ON THE AFFECTED VOLTAGE SELECTOR BOARD WAS PERFORMED. THE TESTS AND INVESTIGATION REVEALED THAT THE DESIGN IS ADEQUATE AND THE NTC "R15" IS WORKING WITHIN ITS SPECIFIED LIMITS. IT WAS CONCLUDED THAT THE DEFECTIVE COMPONENT NTC "R15" IS AN ISOLATED, RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION THE IDENTIFIED ROOT CAUSE IS A DEFECTIVE COMPONENT, I.E. NTC "R15" ON THE VOLTAGE SELECTOR BOARD. THE ISSUE IS CONSIDERED AN ISOLATED, RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE WITH NO INDICATIONS OF AN ADVERSE TREND.
AN INVESTIGATION OF THE INCIDENT IS CURRENTLY UNDERWAY AND A FOLLOW-UP WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE FOLLOWING INVESTIGATION.
LEICA MICROSYSTEMS ((B)(6)) (B)(4) RECEIVED A COMPLAINT FROM (B)(6) STATING THAT THE ILLUMINATION OF THE M530 OHX SHUT OFF DURING SURGERY. THERE WAS A LAMP DEFECTIVE ERROR AND A BURN SMELL COMING OUT OF THE UNIT. THE DEVICE STILL CONTINUED TO WORK AFTER RESTART. THERE WAS NO PATIENT INJURY REPORTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235561 | M530 OHX | SURGIAL MICROSCOPE | EPT | LEICA MICROSYSTEMS (SCHWEIZ) AG | M530 OHX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |