FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PLUS PROFILE

MDR report key: 11336078 · Received February 16, 2021

Report

Report Number
1645337-2021-01597
Event Type
Injury
Date Received
February 16, 2021
Date of Event
January 26, 2021
Report Date
January 26, 2021
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001454
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 04-MAR-2021, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON 08-MAR-2021, MENTOR RECEIVED ADDITIONAL INFORMATION ABOUT THE EVENT. THE PATIENT HAD BOTH OF HER BREAST PROSTHESES REMOVED AND THEY WERE REPLACED WITH MENTOR SMOOTH ROUND HIGH PROFILE 560CC SALINE BREAST PROSTHESES. ON 16-MAR-2021, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED DEFLATION IN THE BREAST IMPLANT. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE HAD A TEAR WITHIN A CREASE/FOLD ON THE ANTERIOR VIEW MEASURING APPROXIMATELY 0.3 CM. THE EVALUATION DETERMINED THAT THE CAUSE OF THE RUPTURE IS CONSISTENT WITH NORMAL WEAR. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN DEFLATION AT A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. A SECOND PRODUCT WAS RECEIVED WITH LOT #6853547. NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE. THEREFORE, NO FURTHER ANALYSIS IS REQUIRED. IT SHOULD BE NOTED THAT AS PART OF MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: BREAST PROSTHESIS DEFLATION. EXPLANTATION MONTH, DAY, AND YEAR: (B)(6) 2021. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR SMOOTH ROUND MODERATE PLUS PROFILE 450CC SALINE BREAST PROSTHESIS THAT DEFLATED AFTER IMPLANTATION. DEFLATION OF THE PATIENT¿S LEFT BREAST PROSTHESIS WAS DIAGNOSED DURING AN OFFICE VISIT. AS A RESULT, THE PATIENT IS SCHEDULED TO UNDERGO EXPLANTATION ON (B)(6) 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230801 MENTOR SMOOTH ROUND MODERATE PLUS PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3502450 6853547 00081317001454

Patients

Seq Age Sex Outcome Treatment
1