FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 11335678 · Received February 16, 2021

Report

Report Number
1221359-2021-00312
Event Type
Malfunction
Date Received
February 16, 2021
Date of Event
January 15, 2021
Report Date
February 16, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 1013573 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING BATCH RECORDS AND QUALITY CONTROL RELEASE TESTING FOR KIT PART NUMBER 191-000 / LOT 1013573 AND TEST BASE PART NUMBER 191-430 / LOT 1013573 WERE REVIEWED. THIS LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS AND FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1013573 SHOWED THAT THE COMPLAINT RATES IS (B)(4) FOR BOTH. ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN THE STATEMENTS MADE WITHIN PACKAGE INSERT RELATED TO % TEST AGREEMENT WITH THE EXPECTED RESULTS BY SAMPLE CONCENTRATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED CONFLICTING RESULTS WITH THE ID NOW COVID-19 ASSAY. THE CUSTOMER REPORTED POSITIVE RESULTS ON A KITTED SWAB WITH THE ID NOW COVID-19 ASSAY PERFORMED (B)(6) 2021 AT 1308. REPEAT TESTING WITH THE COVID-19 ASSAY USING A DIFFERENT ID NOW MACHINE PROVIDED NEGATIVE RESULTS. SAMPLE TYPE AND USE OF VIRAL TRANSPORT MEDIA WERE NOT PROVIDED. NO CONFIRMATION TESTING WAS PERFORMED. NO PATIENT INFORMATION, INCLUDING SYMPTOMS, TREATMENT, AND OUTCOME, WAS PROVIDED. POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2 RNA. PER THE ID NOW COVID-19 PRODUCT INSERT, PRECAUTIONS INCLUDE: DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN THE CLEANING SECTION OF THE INSTRUMENT USER MANUAL. REFER TO SECTION 1.6, MAINTENANCE & CLEANING, FOR FURTHER INFORMATION. NEGATIVE RESULTS SHOULD BE TREATED AS PRESUMPTIVE AND, IF INCONSISTENT WITH CLINICAL SIGNS AND SYMPTOMS OR NECESSARY FOR PATIENT MANAGEMENT, SHOULD BE TESTED WITH DIFFERENT AUTHORIZED OR CLEARED MOLECULAR TESTS. NEGATIVE RESULTS DO NOT PRECLUDE SARS-COV-2 INFECTION AND SHOULD NOT BE USED AS THE SOLE BASIS FOR PATIENT MANAGEMENT DECISIONS. NEGATIVE RESULTS SHOULD BE CONSIDERED IN THE CONTEXT OF A PATIENT'S RECENT EXPOSURES, HISTORY AND THE PRESENCE OF CLINICAL SIGNS AND SYMPTOMS CONSISTENT WITH COVID-19. UNEXPLAINED CONFLICTING RESULTS SHALL BE REPORTED AS IT IS UNKNOWN WHICH RESULT IS CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
227684 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1013573 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 96 YR