FDA Adverse Event Malfunction Summary report: N

ENDOWRIST PROGRASP INSTRUMENT

MDR report key: 1133564 · Received August 22, 2008

Report

Report Number
2955842-2008-01231
Event Type
Malfunction
Date Received
August 22, 2008
Report Date
August 22, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THAT THE DISTAL END OF THE MAIN TUBE HAS A SCRATCHED SECTION WITH LIGHT MATERIAL REMOVED. THE SCRATCHES ARE NOT ALIGNED WITH THE TUBE AXIS AND WERE MOST LIKELY FROM INSTRUMENT COLLISIONS. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SHAFT OF THE ENDOWRIST PROGRASP INSTRUMENT IS ROUGH AND SCRATCHED. NO ADDITIONAL INFO WAS PROVIDED. NO PT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOWRIST PROGRASP INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL, INC. 420093-07 2109071 468

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| DA VINCI S SURGICAL SYSTEM