FDA Adverse Event Malfunction Summary report: N

CATALYS SYSTEM

MDR report key: 11335470 · Received February 16, 2021

Report

Report Number
3006695864-2021-07163
Event Type
Malfunction
Date Received
February 16, 2021
Date of Event
January 22, 2021
Report Date
February 16, 2021
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
OOE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

PHONE : (B)(4). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT; LASER CORRECTION SYSTEM, MODEL CATALYS-I, THAT HAS A SIMILAR PRODUCT, CATALYS-U WHICH IS DISTRIBUTED IN THE UNITED STATES UNDER 510(K) # K200056. A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, SERVICE HISTORY, AND TRENDING WAS REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. A LABELING REVIEW WAS CONDUCTED; THE OPERATOR MANUAL FOR THE SYSTEM WAS REVIEWED AND FOUND TO INCLUDE ADEQUATE INSTRUCTIONS FOR USE, WARNINGS AND OPERATIONAL ERRORS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THE SYSTEM SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. JOHNSON & JOHNSON SURGICAL VISION WILL CONTINUE TO MONITOR THIS TYPE OF COMPLAINTS. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SURGERY MACHINE SHOWED ERROR 03-003 (DISPOSABLE LENS DETECTION FAILURE) INTERRUPTING THE LASER TREATMENT, CAPSULOTOMY CORRECTLY DONE BUT LENS FRAGMENTATION WAS ABORTED. THE SURGEON COMPLETED THE PROCEDURE MANUALLY. NO FURTHER ISSUES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
229609 CATALYS SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE AMO MANUFACTURING USA, LLC CATALYS-I

Patients

Seq Age Sex Outcome Treatment
1