FDA Adverse Event Malfunction Summary report: N

PROLIEVE THERMODILITATION KIT

MDR report key: 1133509 · Received August 22, 2008

Report

Report Number
3005099803-2008-03746
Event Type
Malfunction
Date Received
August 22, 2008
Date of Event
July 24, 2008
Report Date
July 24, 2008
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
MEQ
PMA / PMN Number
P030006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS IS NOT AVAILABLE, AND WE ARE NOT ABLE TO DETERMINE THE RELATIONSHIP BETWEEN THIS DEVICE AND THE CAUSE FOR THIS EVENT.

Description of Event or Problem · 1

IN 2008, IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A PROLIEVE THERMODILITATION KIT WAS USED DURING A BENIGN PROSTATIC HYPERPLASIA (BPH) PROCEDURE. ACCORDING TO THE COMPLAINANT, AT FIFTEEN MINUTES INTO THE PROCEDURE, THE LOW WATER SENSOR WENT OFF, AND THE HEAT EXCHANGER CARTRIDGE WAS REFILLED. THE SENSOR WENT OFF AGAIN, AND IT WAS DETERMINED THAT THE CATHETER HAD A PIN HOLE LEAK. ANOTHER PROLIEVE THERMODILITATION CATHETER WAS USED, AND THE CASE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROLIEVE THERMODILITATION KIT MEQ BOSTON SCIENTIFIC CORPORATION M0068808022 0000606600

Patients

Seq Age Sex Outcome Treatment
1 UNK