GEMINI STONE RETRIEVAL PAIRED WIRE / HELICAL
Report
- Report Number
- 3005099803-2008-03635
- Event Type
- Malfunction
- Date Received
- August 22, 2008
- Report Date
- May 22, 2008
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- FFL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
NO RESIDUE WAS OBSERVED ON THE RETURNED DEVICE, INDICATING IT HAS NOT BEEN USED. A VISUAL INSPECTION OF THE DEVICE FOUND THE SHEATH IS BUCKLED AND SEPARATED AT THE HANDLE. IT WAS ALSO FOUND THAT THE DISTAL 3 CM OF THE DEVICE HAS BEEN CUT OFF FROM THE REST OF THE DEVICE. THE MAJORITY OF THE BASKET AND A SMALL SECTION OF THE SHEATH ARE SEPARATED FROM THE DEVICE. THE CLEAN AND EVEN CUTS THROUGH THE SHEATH AND BASKET WIRES INDICATE THE END OF THE DEVICE WAS VERY LIKELY CUT OFF BY THE USER, AFTER THE SHEATH DAMAGE OCCURRED. CUSTOMER COMPLAINT CONFIRMED. THE RETURNED DEVICE DID NOT FUNCTION FOR THE USER DUE TO SHEATH DAMAGE. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED 2 DEVICES WERE SCRAPPED DURING MFG FOR BEING NONFUNCTIONAL. A LOT HISTORY SEARCH WAS PERFORMED AND FOUND NO SIMILAR COMPLAINTS HAVE BEEN REPORTED FROM THIS LOT. THE MAY 2008 PRODUCT FAMILY TRENDING CHART WAS REVIEWED AND DOES NOT SHOW A NEGATIVE TREND.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON 05/22/2008 THAT ON AN UNK DATE, WHILE UNPACKING A MINI GEMINI BASKET, THE DEVICE WAS NOTICED TO BE DAMAGED. ALTHOUGH THE DEVICE WAS RETURNED WITH THE BASKET DETACHED, THE COMPLAINANT CONFIRMED THAT THE BASKET WAS NOT DETACHED WHEN SHIPPED FROM THEIR FACILITY. THE COMPLAINANT INDICATED THAT THE DEVICE WAS NOT USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GEMINI STONE RETRIEVAL PAIRED WIRE / HELICAL | FFL | BOSTON SCIENTIFIC CORPORATION | M0063303060 | 11135814 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |