MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2008-01239
- Event Type
- Malfunction
- Date Received
- August 22, 2008
- Report Date
- August 22, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE TUBE EXTENSION IS DISLODGED FROM THE INSTRUMENT MAIN TUBE, WITH THE THIN WALLED SECTION OF THE TUBE EXTENSION BROKEN CIRCUMFERENTIALLY AT THE INTERFACE. NO PIECES ARE MISSING. ENGINEERING ALSO OBSERVED CHARRING ON THE TUBE EXTENSION PIECES, INDICATING THAT AN ARCING EVENT OCCURRED. ENGINEERING ALSO DISCOVERED A CHARRED .120" X .060" OBLONG SECTION ON THE DISTAL END OF THE MAIN TUBE, LOCATED .5" ABOVE THE TUBE EXTENSION. THE CHARRING INDICATES THAT A SECOND ARCING EVENT OCCURRED. THE INSTRUMENT WAS DISASSEMBLED AND A SMALL CHAR WAS NOTICED ON THE TUBE EXTENSION SEAL. ENGINEERING ALSO FOUND A CRACK IN THE TUBE THAT STARTS AT ONE SLOT AND ENDS AT THE CHARRED SECTION. IT WAS CONCLUDED THAT THE TUBE EXTENSION BREAKAGE MAY HAVE PROPAGATED THE TUBE CRACK AT THE SLOT WHICH LIKELY CREATED A PATH FOR ARCING TO OCCUR AND HIGH GENERATOR SETTINGS MAY HAVE ALSO CONTRIBUTED TO THE DAMAGE.
IT WAS REPORTED THAT AFTER A DA VINCI S TOTAL HYSTERECTOMY SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT SNAPPED, HOWEVER, NOTHING FELL INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | GEI | INTUITIVE SURGICAL, INC. | 420179-08 | 0804281 237 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| DA VINCI S SURGICAL SYSTEM| ELECTROSURGICAL UNIT |