ENDO CLIP III
Report
- Report Number
- 9612501-2021-00253
- Event Type
- Malfunction
- Date Received
- February 16, 2021
- Date of Event
- January 22, 2021
- Report Date
- March 31, 2021
- Manufacturer
- DAVIS & GECK CARIBE LTD
- Product Code
- FZP
- UDI-DI
- 10884521057852
- PMA / PMN Number
- K100242
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE EVALUATION FOUND NO POTENTIALLY CONTRIBUTING FACTORS, AND THE SAMPLE MET ALL RELATED SPECIFICATIONS. IT WAS REPORTED THAT THE CLIPS DID NOT CLOSE COMPLETELY AND WERE MALFORMED. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE IDENTIFIED BECAUSE NO RELATED PROBLEM WAS DETECTED WITH THE DEVICE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: FIRING A CLIP OVER ANOTHER CLIP OR OTHER OBSTRUCTIONS MAY RESULT IN BLEEDING AND/OR LEAKAGE AND MAY DAMAGE THE INSTRUMENT JAWS. ALSO, AVOID EXCESSIVE TWISTING OF THE JAWS OR TISSUE MANIPULATION WHEN FIRING THE INSTRUMENT. DEFLECTING THE JAWS AND/OR SHAFT DURING FIRING MAY RESULT IN AN IMPROPERLY FORMED CLIP AND POSSIBLE BLEEDING AND/OR LEAKAGE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ACCORDING TO THE REPORTER, AT THE FIRST FIRING DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE CLIPS APPEARED TO BE NOT FULLY CLOSED. THE CLIPS AT THE TIP WERE CLOSED, BUT THERE WAS A GAP IN THE PROXIMAL PART. AT THE SECOND FIRING, THE CLIPS WERE APPLIED JUST NEXT TO THE FIRST FIRING AND CLIPS FORMED WITHOUT ANY PROBLEMS. WHEN FIRING IN THE AIR SEVERAL TIMES OUTSIDE THE BODY, THERE WERE TIMES WHEN THE CLIPS WERE ACCURATELY CLIPPED AND TIMES WHEN THEY BECAME UNFORMED. THE STATE OF U SHAPE AND THE ONLY TIP WAS CLOSED AND THE CLIPS CAME OUT SO THE USE OF THE CLIP APPLIER WAS STOPPED AND NEW PRODUCT WAS TAKEN OUT. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232364 | ENDO CLIP III | CLIP, IMPLANTABLE | FZP | DAVIS & GECK CARIBE LTD | 176630 | J0C0402Y | 10884521057852 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |