ARCHITECT C8000 SYSTEM
Report
- Report Number
- 1628664-2008-00201
- Event Type
- Malfunction
- Date Received
- August 19, 2008
- Date of Event
- July 25, 2008
- Report Date
- July 25, 2008
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JJE
- PMA / PMN Number
- K980367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
A REVIEW OF THE COMPLAINT TEXT INDICATES THE ARCHITECT C8000 GENERATED AN ERRATIC CREATININE RESULTS FOR ONE PATIENT IN 2008. THE HISTORICAL DATA OF THIS PATIENT WAS AN AVERAGE OF 8 FOR CREATININE. THE PATIENT SAMPLE GENERATED AN INITIAL CREATININE RESULT OF 5.3 AND GENERATED THE RESULTS OF 8.9 AND 9.2, RESPECTIVELY, WHEN IT WAS REPEATED. A REPRODUCIBILITY TEST WAS PERFORMED AND THE CONTROLS PASSED THE SPECIFICATIONS. THE OPERATOR STATED THE QUARTERLY MAINTENANCE WAS COMPLETED IN THE MONTH PRIOR. THE OPERATOR CHECKED THE R1 PROBE, R2 PROBE, WASH UNIT PROBE, AND THE AGITATORS ON THE INSTRUMENT. THE SAMPLE PROBE WAS THEN REPLACED. ONCE THE SAMPLE PROBE WAS PLACED ONTO THE INSTRUMENT, ERROR CODE 3375 (UNABLE TO PROCESS TEST, ASPIRATION ERROR OCCURRED) WAS GENERATED. THE OPERATOR THEN ADJUSTED THE PLACEMENT AND ROUTING OF THE SAMPLE PROBE TO REALIZE THE CALIBRATION TOOL WAS NOT REMOVED FROM THE CAROUSEL FROM THE MAINTENANCE PROCEDURE. THE CALIBRATIONS PASSED AND THE OPERATOR HAS NOT OBSERVED ANY OTHER ISSUES WITH ERRATIC RESULTS SINCE THE REPLACEMENT OF THE SAMPLE PROBE. IN THE ARCHITECT SYSTEM OPERATIONS MANUAL ((PN 201837-104) JUNE, 2007): SECTION 7, OPERATIONAL PRECAUTIONS AND LIMITATIONS, IT STATESLIMITATIONS OF RESULT INTERPRETATIONASSAY RESULTS MUST BE USED WITH OTHER CLINICAL DATA, FOR EXAMPLE, SYMPTOMS, OTHER TEST RESULTS, PATIENT HISTORY, CLINICAL IMPRESSIONS, INFORMATION AVAILABLE FROM CLINICAL EVALUATION, AND OTHER DIAGNOSTIC PROCEDURES. ALL DATA MUST BE CONSIDERED FOR PATIENT CARE MANAGEMENT.IF ASSAY RESULTS ARE INCONSISTENT WITH CLINICAL EVIDENCE, ADDITIONAL TESTING IS SUGGESTED TO CONFIRM THE RESULT.THE ARCHITECT SYSTEM HAS BEEN VALIDATED FOR ITS INTENDED USE. HOWEVER, ERRORS CAN OCCUR DUE TO POTENTIAL OPERATOR ERRORS AND ARCHITECT SYSTEM TECHNOLOGY LIMITATIONS.IN SECTION 9: SERVICE AND MAINTENANCE: COMPONENT REPLACEMENT PROVIDES THOROUGH INSTRUCTIONS ON THE PROCEDURE TO REMOVE, REPLACE, AND VERIFY A SAMPLE PROBE. IN SECTION 10: TROUBLESHOOTING AND DIAGNOSTICS UNDER OBSERVED PROBLEMS LIST MULTIPLE PROBABLE CAUSES AND RESOLUTIONS RELATED TO THE CUSTOMERS ISSUE.IN THE CLINICAL CHEMISTRY CREATININE REAGENT PACKAGE INSERT (30-3916/R7), LITERATURE IS PROVIDED IN DESCRIBING SUITABLE SPECIMENS, RESULTS, AND LIMITATIONS OF THE PROCEDURE.THE ISSUE WAS MOST LIKELY CAUSED BY AN OBSTRUCTED SAMPLE PROBE. AFTER THE REPLACEMENT OF THE SAMPLE PROBE, THE INSTRUMENT WAS RUNNING WITHIN SPECIFICATIONS. THIS IS THE FINAL REPORT.
THE ACCOUNT STATED THAT THE ARCHITECT C8000 ANALYZER IS GENERATING ERRATIC CREATININE RESULTS FOR ONE PATIENT. THE INITIAL RESULT WAS 5.3, THE SAMPLE WAS THEN REPEATED TWICE AND THE RESULTS WERE 5.9 AND 9.2. THE PATIENT WAS KNOWN TO HAVE A CREATININE RESULT OF 8.0. (UNITS OF MEASUREMENT WAS NOT PROVIDED). THE QC WAS WITHIN RANGE. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT C8000 SYSTEM | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |