FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1132729 · Received August 21, 2008

Report

Report Number
1644487-2008-01979
Event Type
Malfunction
Date Received
August 21, 2008
Date of Event
July 30, 2008
Report Date
July 30, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PROGRAMMING SYSTEM WAS PRODUCING COMMUNICATION ERRORS WHEN TRYING TO COMMUNICATE WITH A DEMO GENERATOR. TROUBLESHOOTING DID NOT RESOLVE THE ISSUE. THE PROGRAMMING WAND WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. PRODUCT ANALYSIS REVEALED THAT THE SERIAL ADAPTER CABLE HAD AN INTERMITTENT CONDUCTOR WHICH WAS PRODUCING THE COMMUNICATION ERRORS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND PROGRAMMING WAND LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1