FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 1132642 · Received August 22, 2008

Report

Report Number
2953200-2008-00721
Event Type
Injury
Date Received
August 22, 2008
Date of Event
July 22, 2008
Report Date
July 25, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVAL RESULTS: (ARTERIAL AND VENOUS OCCLUSION). LACK OF INFO (NO OPERATIVE REPORTS OR FILMS WERE RECEIVED). (REQUIRED SECONDARY INTERVENTION).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYS WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM 20 DAYS AGO. ANEURYSM AND VESSEL MORPHOLOGY WERE NOT REPORTED. IT WAS REPORTED THAT THE PT PRESENTED WITH DIMINISHED FLOW THROUGH THE NON STENTED RING AREA OF THE BIFURCATED STENT GRAFT. A PALMAZ STENT WAS IMPLANTED AT THE LOCATION OF THE NON STENTED RING AREA AND SUCCESSFULLY RESOLVED THE OCCLUSION. NO ADD'L CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI