FDA Adverse Event Injury Summary report: N

GLV EXAM NITRILE ESTEEM COMFORT LG

MDR report key: 11323849 · Received February 12, 2021

Report

Report Number
1423537-2021-00594
Event Type
Injury
Date Received
February 12, 2021
Date of Event
January 19, 2021
Report Date
March 4, 2021
Manufacturer
SGP01 SINGAPORE
Product Code
LZC
UDI-DI
02088538016941
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NOTE ON THE ORIGINAL SUBMISSION THE PRODUCT CODE LCZ WAS USED IN ERROR. THE CORRECT PRODUCT CODE IS LZC. AT THE TIME OF INVESTIGATION, NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER. WITHOUT THE LOT NUMBER, WE ARE UNABLE TO TRACE THE MANUFACTURER OR REVIEW THE DEVICE HISTORY RECORDS FOR FURTHER INVESTIGATION. NO SAMPLE WAS PROVIDED AT THE TIME OF THIS INVESTIGATION. BASED ON INVESTIGATION THE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO NO LOT NUMBER OR SAMPLE BEING PROVIDED. NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME, DUE TO ROOT CAUSE NOT BEING IDENTIFIED. WE WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

AT THE TIME OF INVESTIGATION, NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER. WITHOUT THE LOT NUMBER, WE ARE UNABLE TO TRACE THE MANUFACTURER OR REVIEW THE DEVICE HISTORY RECORDS FOR FURTHER INVESTIGATION. NO SAMPLE WAS PROVIDED AT THE TIME OF THIS INVESTIGATION. BASED ON INVESTIGATION THE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO NO LOT NUMBER OR SAMPLE BEING PROVIDED. NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME, DUE TO ROOT CAUSE NOT BEING IDENTIFIED. WE WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

BASED ON INFORMATION RECEIVED FROM THE END-USER, SHE REPORTEDLY HAD HAND AND ARM SWELLING, INFLAMMATION, FACIAL SWELLING, FULL BODY ITCHING. REPORTEDLY SHE HAS SEEN DOCTORS SEVERAL TIMES. THE (B)(6) CLINIC DERMATOLOGIST IN (B)(6) IS CALLING THIS A TYPE 4 ALLERGY. SHE REPORTEDLY HAD PATCH TESTING DONE WHICH SHOWS ALLERGIC TO A RUBBER PRODUCT. SHE WAS ADVISED TO USE SKIN LOTION AND VANICREAM AND HAS BEEN GIVEN A STEROID FOR SEVERE SWELLINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221558 GLV EXAM NITRILE ESTEEM COMFORT LG PATIENT EXAMINATION GLOVE, SPECIALTY LZC SGP01 SINGAPORE 88ES04L UNKNOWN 02088538016941

Patients

Seq Age Sex Outcome Treatment
1 Other