FDA Adverse Event Malfunction Summary report: N

ADVANTA

MDR report key: 1132293 · Received August 22, 2008

Report

Report Number
1824206-2008-03126
Event Type
Malfunction
Date Received
August 22, 2008
Date of Event
August 4, 2008
Report Date
August 4, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ACCOUNT ALLEGED INTERMITTENT BED OPERATION. NO INJURY REPORTED. UPON INVESTIGATION, THE POWER CORD WAS FOUND TO HAVE AN OPEN ON ONE OF THE WIRES CAUSING THE BED TO HAVE NO POWER. THE TECH REPLACED THE POWER CORD TO RESOLVE THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1600 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK