FDA Adverse Event Malfunction Summary report: N

ATTUNE RP TIB BASE SZ 3 CEM

MDR report key: 11320906 · Received February 12, 2021

Report

Report Number
1818910-2021-02866
Event Type
Malfunction
Date Received
February 12, 2021
Date of Event
August 20, 2020
Report Date
February 9, 2021
Manufacturer
DEPUY IRELAND - 9616671
Product Code
NJL
UDI-DI
10603295492153
PMA / PMN Number
P830055
Removal / Correction Number
1818910-02/22/2021-002-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT#: (B)(4). INVESTIGATION SUMMARY: UPDATE 24-FEB-2021: THE INVESTIGATION WAS RE-OPENED DUE TO NR-0156854. EXAMINATION OF THE RETURNED DEVICE CONFIRMED THE PACKAGING WAS DAMAGED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: THE DHR REVIEW IDENTIFIED THE FOLLOWING: PRODUCT CODE 150680003, WORK ORDER: (B)(4) WAS MANUFACTURED ON THE 08 JULY 2020. 20 PARTS WERE MANUFACTURED PER SPECIFICATION AND ALL RAW MATERIAL MET SPECIFICATION. THERE WERE NO NON-CONFORMANCES (NC) FOUND TO BE ASSOCIATED WITH WORK ORDER: (B)(4). THERE WERE NO SCRAP PARTS FOUND TO BE ASSOCIATED WITH WORK ORDER: (B)(4). THERE WAS NO REPROCESSING ASSOCIATED WITH WORK ORDER: (B)(4).

Additional Manufacturer Narrative · 0

H7 AND H9: RECALL NUMBER PROVIDED.

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN NURSE WENT TO OPEN THE SIZE 3 RP TIBIA, THE PACKAGING WAS HARD PLASTIC AND LOOKED LIKE IT WAS EXPOSED TO HEAT AS IT WAS PRESSURIZED. SURGEON ASKED FOR A NEW IMPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
222301 ATTUNE RP TIB BASE SZ 3 CEM KNEE TIBIAL TRAY NJL DEPUY IRELAND - 9616671 1506-80-003 9557440 10603295492153

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female