FDA Adverse Event
Malfunction
Summary report: N
VDW GOLD/SILVER CONTRA ANGLE
MDR report key: 11319769
·
Received February 11, 2021
Report
- Report Number
- 9611053-2021-00010
- Event Type
- Malfunction
- Date Received
- February 11, 2021
- Report Date
- February 11, 2021
- Manufacturer
- DENTSPLY VDW GMBH
- Product Code
- EGS
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. BATTERY (VOLTAGE COLLAPSES UNDER LOAD) DEFECTIVE, REPLACED. CONTRA-ANGLE HANDPIECE 92038 (2016) (NO LONGER HOLDS THE FILE) DEFECTIVE, REPLACED WITH 100487 (2017). MICRO MOTOR GMR1657836 AND FOOT CONTROL 122704 WITHOUT ERRORS. FUNCTION TEST AND TEST MEASUREMENTS WITHOUT ERRORS.
Description of Event or Problem · 1
IN THIS EVENT IT WAS REPORTED THAT A CONTRA ANGLE 6:1 FOR VDW.GOLD/SILVER WOULD NOT HOLD FILES; NO INJURY RESULTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 215917 | VDW GOLD/SILVER CONTRA ANGLE | HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL | EGS | DENTSPLY VDW GMBH | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |