FDA Adverse Event Malfunction Summary report: N

CNS-6801A

MDR report key: 11319456 · Received February 11, 2021

Report

Report Number
8030229-2021-00071
Event Type
Malfunction
Date Received
February 11, 2021
Date of Event
January 12, 2021
Report Date
May 4, 2021
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
UDI-DI
04931921131640
PMA / PMN Number
K102376
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

DETAILS OF COMPLAINT: ON 01/12/2021, THE CUSTOMER REPORTED THAT THE CENTRAL NURSE'S STATION (CNS) HAD AN HDD PORT 1 AND HDD PORT2 ERROR FLASHING ON THE DISPLAY SCREEN WHILE MONITORING PATIENTS. SERVICE REQUESTED / PERFORMED: TROUBLESHOOTING. INVESTIGATION SUMMARY: THE OVERALL RISK SCORE IS LOW. THE REPORTED DEVICE WAS NOT RETURNED FOR EVALUATION. NK TECH SUPPORT STATED THAT THIS ISSUE WAS A RESULT OF THE HDD'S OPERATING FOR MORE THAN 20,000 HOURS. NK TS WALKED THE CUSTOMER THROUGH HOW TO REMOVE THESE ERROR MESSAGES PER TN-1270. AS THIS ISSUE IS NOT A RESULT OF A DEFICIENCY OR NON-CONFORMANCE OF AN NK DEVICE, A ROOT CAUSE ANALYSIS AND CAPA ARE NOT REQUIRED. THE FOLLOWING FIELDS CONTAIN NO INFORMATION (NI), AS ATTEMPTS TO OBTAIN INFORMATION WERE MADE, BUT NOT PROVIDED. ATTEMPT # 1: 02/11/2021 EMAILED THE CUSTOMER VIA MICROSOFT OUTLOOK FOR ALL THE INFORMATION : NO REPLY WAS RECEIVED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED OF HARD DRIVE (HDD) PORT 1 AND HARD DRIVE (HDD) PORT2 ERROR FLASHING ON CENTRAL NURSING STATION (CNS), THEY SAW HARD DRIVE (HDD) PORT 1 AND HARD DRIVE (HDD) PORT2 ERROR FLASHING ON THE CENTRAL NURSING STATION(CNS) AND THERE WAS NO OTHER ISSUES WITH THE CENTRAL NURSE'S STATION (CNS) ONLY THAT THESE ERRORS WERE DISPLAYED ON THE CENTRAL NURSE'S STATION (CNS) SOFTWARE. NO PATIENT HARM WAS REPORTED.

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED OF HARD DRIVE (HDD) PORT 1 AND HARD DRIVE (HDD) PORT2 ERROR FLASHING ON CENTRAL NURSING STATION (CNS). THEY SAW HARD DRIVE (HDD) PORT 1 AND HARD DRIVE (HDD) PORT2 ERROR FLASHING ON THE CENTRAL NURSING STATION(CNS) AND THERE WAS NO OTHER ISSUES WITH THE CENTRAL NURSE'S STATION (CNS) ONLY THAT THESE ERRORS WERE DISPLAYED ON THE CENTRAL NURSE'S STATION (CNS) SOFTWARE. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED OF HARD DRIVE (HDD) PORT 1 AND HARD DRIVE (HDD) PORT2 ERROR FLASHING ON CENTRAL NURSING STATION (CNS), THEY SAW HARD DRIVE (HDD) PORT 1 AND HARD DRIVE (HDD) PORT2 ERROR FLASHING ON THE CENTRAL NURSING STATION(CNS) AND THERE WAS NO OTHER ISSUES WITH THE CENTRAL NURSE'S STATION (CNS) ONLY THAT THESE ERRORS WERE DISPLAYED ON THE CENTRAL NURSE'S STATION (CNS) SOFTWARE. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
215813 CNS-6801A CENTRAL MONITOR SYSTEM MHX NIHON KOHDEN CORPORATION CNS-6801A NA 04931921131640

Patients

Seq Age Sex Outcome Treatment
1 NI.| NI.