FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 11319153 · Received February 11, 2021

Report

Report Number
1645337-2021-01486
Event Type
Injury
Date Received
February 11, 2021
Date of Event
January 26, 2021
Report Date
January 26, 2021
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317001676
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2021, IT WAS REPORTED TO MENTOR THAT THE AFFECTED DEVICE WAS MENTOR MEMORYGEL BREAST IMPLANT 500CC, CATALOG NUMBER 3505001BC, LOT NUMBER 5890302. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 5890302 NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 3/17/2021 , UPON VISUAL EVALUATION OF THE IMAGE PROVIDED IN THE COMPLAINT ON 2/1/2021, THE IMPLANT WAS OBSERVED COMPLETELY RUPTURED. AS THE PRODUCT INVOLVED IN THIS COMPLAINT WAS NOT RECEIVED, THE DEVICE COULDN´T BE ANALYZED ACCORDING TO OUR PROCEDURES. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE NUMBER 5890302 AND NO NON-CONFORMANCES WERE IDENTIFIED. AS PART OF MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH A MENTOR SMOOTH ROUND HIGH PROFILE 460CC AND SUFFERED FROM A BILATERAL RUPTURE. AS A RESULT, THE PATIENT HAD UNDERGONE REMOVAL ON (B)(6) 2021. THIS MEDWATCH IS FOR THE RIGHT BREAST IMPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
215884 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 3503460 5890302 00081317001676

Patients

Seq Age Sex Outcome Treatment
1 32 YR Required Intervention