FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK 50ML CONCENTRIC LUER LOCK SYRINGE

MDR report key: 11312266 · Received February 11, 2021

Report

Report Number
3003152976-2021-00075
Event Type
Malfunction
Date Received
February 11, 2021
Date of Event
January 19, 2021
Report Date
April 6, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: (B)(6) 2021.INVESTIGATION SUMMARY:TWO USED SAMPLES AND THREE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PRODUCT WAS VISUALLY INSPECTED AND A LEAKAGE PAST THE STOPPER RIBS WAS OBSERVED. THERE ARE NO DEFECTS OR DAMAGE WAS NOTED ON ANY OF THE SAMPLES OR COMPONENTS AND THE STOPPERS WERE VERIFIED TO BE PROPERLY ASSEMBLED ONTO THE PLUNGER ROD. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2010022, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES, INCLUDING LEAKAGE TESTING AND TIGHTNESS TESTING TO VERIFY THE SEAL OF THE STOPPER. THE RETURNED SAMPLES UNDERWENT THESE TESTS AND IN ALL CASES THE PRODUCT MET REQUIRED SPECIFICATION. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE MEDICATION PROPOFOL LEAKED ALONG THE PESTLE WHEN THE LIQUID IS PULLED UP.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE MEDICATION PROPOFOL LEAKED ALONG THE PESTLE WHEN THE LIQUID IS PULLED UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
210717 BD PLASTIPAK 50ML CONCENTRIC LUER LOCK SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. 2010022

Patients

Seq Age Sex Outcome Treatment
1