FDA Adverse Event Malfunction Summary report: N

BD SOLOMED SYRINGE

MDR report key: 11312154 · Received February 11, 2021

Report

Report Number
3003916417-2021-00027
Event Type
Malfunction
Date Received
February 11, 2021
Date of Event
January 18, 2021
Report Date
February 11, 2021
Manufacturer
BECTON DICKINSON IND. CIRURGICAS LTDA
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION: IT WAS PERFORMED THE DHR, QUALITY NOTIFICATIONS AND MAINTENANCE RECORDS WHERE NO RECORDS POTENTIALLY RELATED TO FAILURE WAS FOUND. THE PHOTO AND SAMPLE PROVIDED WAS ANALYZED AND IT WAS POSSIBLE TO OBSERVE BARREL CRACKED. THE POTENTIAL CAUSE FOR THIS IS A SYRINGE ASSEMBLY FAILURE THAT CAUSED THE DAMAGE. CAPA #1035416 WAS INITIATED. THIS IS THE 2ND RELATED COMPLAINT FOR BARREL CRACKED ON THE PROVIDED LOT NUMBER 9190268. H3 OTHER TEXT: SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 6 BD SOLOMED¿ SYRINGES WERE CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: "CUSTOMER REPORTS THAT WHEN APPLYING THE PRODUCT, IT WAS NOTICED THAT THE SYRINGE WAS CRACKED, LEADING TO THE WASTE OF THE PRODUCT."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT 6 BD SOLOMED¿ SYRINGES WERE CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "CUSTOMER REPORTS THAT WHEN APPLYING THE PRODUCT, IT WAS NOTICED THAT THE SYRINGE WAS CRACKED, LEADING TO THE WASTE OF THE PRODUCT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213768 BD SOLOMED SYRINGE PISTON SYRINGE FMF BECTON DICKINSON IND. CIRURGICAS LTDA 9190268

Patients

Seq Age Sex Outcome Treatment
1