FDA Adverse Event Malfunction Summary report: N

BD NEXIVA KIT W/ Q-SYTE 22GA 1.0IN

MDR report key: 11312047 · Received February 10, 2021

Report

Report Number
9610847-2021-00062
Event Type
Malfunction
Date Received
February 10, 2021
Date of Event
January 14, 2021
Report Date
March 5, 2021
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A COMPLAINT OF FOREIGN MATTER WAS RECEIVED FROM THE CUSTOMER. A SAMPLE WAS RETURNED TO AID IN THE INVESTIGATION. THROUGH VISUAL INSPECTION FOREIGN MATTER WAS CONFIRMED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED LOT NUMBER 2011252. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. A ROOT CAUSE COULD NOT BE DETERMINED. A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT REPORTED WITH THE DEFECT/CONDITION OF FOREIGN MATTER WITH LOT #2011252 REGARDING ITEM #383652.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD NEXIVA¿ KIT W/ Q-SYTE¿ 22GA 1.0IN EXPERIENCED A DAMAGED OR OPEN UNIT PACKAGE/SEAL WHERE STERILITY WAS COMPROMISED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HAIR WAS FOUND IN A PACKAGE OF Q-SYTE (385100), WHICH IS A COMPONENT OF NEXIVA (383652). LOT # OF Q-SYTE (385100) IS REPORTED TO BE 2011252.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE EXPIRATION DATE: UNKNOWN A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD NEXIVA¿ KIT W/ Q-SYTE¿ 22GA 1.0IN EXPERIENCED A DAMAGED OR OPEN UNIT PACKAGE/SEAL WHERE STERILITY WAS COMPROMISED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HAIR WAS FOUND IN A PACKAGE OF Q-SYTE (385100), WHICH IS A COMPONENT OF NEXIVA (383652). LOT # OF Q-SYTE (385100) IS REPORTED TO BE 2011252.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
206916 BD NEXIVA KIT W/ Q-SYTE 22GA 1.0IN INTERVASCULAR ADMINISTRATION SET FPA BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 2011252

Patients

Seq Age Sex Outcome Treatment
1