FDA Adverse Event Malfunction Summary report: N

SYRINGE 10CC S/T WOS STERILE WATER BNS

MDR report key: 11311496 · Received February 10, 2021

Report

Report Number
1213809-2021-00067
Event Type
Malfunction
Date Received
February 10, 2021
Date of Event
December 11, 2020
Report Date
February 16, 2021
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY TWO PHOTOS OF A LOOSE 10ML STERILE WATER SYRINGE WERE RECEIVED AND EVALUATED. IT WAS OBSERVED THE PLUNGER ROD WAS IN THE BOTTOMED-OUT POSITION AND A PORTION THE STOPPER WAS WEDGED BETWEEN THE PLUNGER ROD AND BARREL WALL. THE JAMMED STOPPER CONDITION WAS REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE JAMMED STOPPER DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 10CC S/T WOS STERILE WATER BNS STOPPER WAS DEFORMED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 304086, BATCH NO: 9170732. IT WAS REPORTED THAT THE GASKET OF THE SYRINGE WAS DEFORMED.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 10CC S/T WOS STERILE WATER BNS STOPPER WAS DEFORMED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 304086, BATCH NO: 9170732. IT WAS REPORTED THAT THE GASKET OF THE SYRINGE WAS DEFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
207979 SYRINGE 10CC S/T WOS STERILE WATER BNS PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9170732

Patients

Seq Age Sex Outcome Treatment
1