BD VACUTAINER URINE ANALYSIS PRESERVATIVE TUBE
Report
- Report Number
- 1917413-2021-00084
- Event Type
- Malfunction
- Date Received
- February 10, 2021
- Date of Event
- January 21, 2021
- Report Date
- March 3, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JTW
- UDI-DI
- 50382903649926
- PMA / PMN Number
- K790366
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: BD RECEIVED FOUR (4) SAMPLES FROM CATALOG 364992, LOT NUMBER 9344900; TWO (2) SAMPLES FROM CATALOG 364957, LOT NUMBER 0120087 AND TWO (2) SAMPLES FROM CATALOG 364992, LOT NUMBER 0167199 FOR INVESTIGATION. THE SAMPLES WERE EVALUATED BY FUNCTIONAL TESTING AND THE INDICATED FAILURE MODE FOR DRAW VOLUME WITH THE INCIDENT LOT WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.
THE FOLLOWING FIELDS HAVE BEEN UPDATED DUE TO ADDITIONAL INFORMATION: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: D.4. MEDICAL DEVICE LOT #: 0167199. D.4. MEDICAL DEVICE EXPIRATION DATE: 11/30/2021. H.4. DEVICE MANUFACTURE DATE: 6/15/2020. D.4. MEDICAL DEVICE LOT #: 9344900. D.4. MEDICAL DEVICE EXPIRATION DATE: 5/31/2021. H.4. DEVICE MANUFACTURE DATE: 12/10/2019.
THE INITIAL REPORTER EXPERIENCED TEN OCCURRENCES OF UNDERFILL OR LOW DRAW WHEN USING THE BD VACUTAINER® URINE ANALYSIS PRESERVATIVE TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT IS REPORTED TUBES ARE UNDER FILLING. RECEIVED CALL FROM LABORATORY MANAGER THAT REPORTS FACILITY NOTICING .5 ML AWAY FROM MINIUM FILL LINE. HE STATES ONLY 10 INSTANCES WITHIN A MONTHS TIME. NO SPECIFIC DATE OF EVENT. FACILITY HAS TO RECOLLECT PATIENT'S URINE WITH NO ISSUES. FACILITY IS SITTING AT 7,000 FEET ELEVATION."
THE INITIAL REPORTER EXPERIENCED TEN OCCURRENCES OF UNDERFILL OR LOW DRAW WHEN USING THE BD VACUTAINER® URINE ANALYSIS PRESERVATIVE TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER; "IT IS REPORTED TUBES ARE UNDER FILLING. RECEIVED CALL FROM LABORATORY MANAGER THAT REPORTS FACILITY NOTICING .5 ML AWAY FROM MINIUM FILL LINE. HE STATES ONLY 10 INSTANCES WITHIN A MONTHS TIME. NO SPECIFIC DATE OF EVENT. FACILITY HAS TO RECOLLECT PATIENT'S URINE WITH NO ISSUES. FACILITY IS SITTING AT 7,000 FEET ELEVATION.".
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
THE INITIAL REPORTER EXPERIENCED TEN OCCURRENCES OF UNDERFILL OR LOW DRAW WHEN USING THE BD VACUTAINER® URINE ANALYSIS PRESERVATIVE TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER; "IT IS REPORTED TUBES ARE UNDER FILLING. RECEIVED CALL FROM LABORATORY MANAGER THAT REPORTS FACILITY NOTICING .5 ML AWAY FROM MINIUM FILL LINE. HE STATES ONLY 10 INSTANCES WITHIN A MONTHS TIME. NO SPECIFIC DATE OF EVENT. FACILITY HAS TO RECOLLECT PATIENT'S URINE WITH NO ISSUES. FACILITY IS SITTING AT 7,000 FEET ELEVATION."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 208671 | BD VACUTAINER URINE ANALYSIS PRESERVATIVE TUBE | SPECIMEN TRANSPORT AND STORAGE CONTAINER | JTW | BECTON, DICKINSON & CO. (BROKEN BOW) | 364992 | SEE H.10 | 50382903649926 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |