FDA Adverse Event Injury Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1130867 · Received August 21, 2008

Report

Report Number
2023826-2008-01094
Event Type
Injury
Date Received
August 21, 2008
Date of Event
July 30, 2008
Report Date
July 30, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL: RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT THERE WERE TEARS IN THE OPTIC AND BOTH HAPTICS WERE TORN OFF FROM THE OPTIC AND MISSING. THERE WAS EVIDENCE OF A CLEAR SURGICAL RESIDUE. CONCLUSION: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CAN NOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015A THREE PIECE COLLAMER LENS AND A HAPTIC TORE UPON INSERTION. THE INCISION WAS ENLARGED SLIGHTLY TO REMOVE THE LENS AND ANOTHER SAME MODEL LENS WAS IMPLANTED. A SUTURE CLOSED THE WOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention INJECTOR MODEL MSI-PM| CARTRIDGE MODEL CQ CARTRIDGE-FP