FDA Adverse Event Other Summary report: N

CORLFO

MDR report key: 1130803 · Received August 14, 2008

Report

Report Number
1419949-2008-00003
Event Type
Other
Date Received
August 14, 2008
Date of Event
July 1, 2008
Report Date
August 12, 2008
Manufacturer
VIASYS MEDSYSTEMS
Product Code
KNT
PMA / PMN Number
K840656
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RETENTION SAMPLES WERE TESTED BY PLACING FEEDING SETS INTO G TUBES. ABOUT HALF OF THE UNITS THAT WERE DIPPED QUICKLY IN ALCOHOL HAD CRACKED AFTER BEING MATED TO A G-TUBE SEVERAL TIMES. UNITS THAT WERE NOT DIPPED IN ALCOHOL REMAINED FAIRLY EASY TO REMOVE AND RE-INSERT INTO THE G-TUBES AND NO SIGNS OF CRACKING WERE SEEN. WITHOUT A SAMPLE INVOLVED WITH THIS PARTICULAR REPORT, WE ARE UNABLE TO DETERMINE THE EXACT CAUSE FOR THE REPORTED CRACKING.

Description of Event or Problem · 1

EVENT DESCRIPTION: PLASTIC TIP OF FEEDING TUBE EXTENSION SET BROKE OFF AND BECAME LODGED IN BABY'S GASTROSTOMY TUBE. SURGEON HAD TO BRING SPECIAL INSTRUMENTATION TO NICU FROM OR TO EXTRACT THE BROKEN PIECE. NO ADDITIONAL INFO PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORLFO ENTERAL Y EXTENSION SET KNT VIASYS MEDSYSTEMS NA 32806

Patients

Seq Age Sex Outcome Treatment
1 3 MO