FDA Adverse Event Malfunction Summary report: N

CATGUT PLAIN 1 (5) 75M

MDR report key: 11307642 · Received February 10, 2021

Report

Report Number
3003639970-2021-00033
Event Type
Malfunction
Date Received
February 10, 2021
Report Date
February 10, 2021
Manufacturer
B. BRAUN SURGICAL, S.A.
Product Code
GAL
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. UNDER PMA/510(K) NUMBER: K991223. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED ONE OPEN AND USED CASSETTE. WE HAVE TESTED THE KNOT PULL TENSILE STRENGTH OF THE THREAD OF THE CASSETTE RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 3.96 KGF IN AVERAGE AND 3.12 KGF IN MINIMUM (EP REQUIREMENTS: 3.80 KGF IN AVERAGE AND 2.00 KGF IN MINIMUM) REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE CASSETTE RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.

Description of Event or Problem · 1

IT WAS REPORTED AN ISSUE WITH CATGUT SUTURE. THE CLIENT REPORTED THAT THE THREAD BREAKS WHEN KNOTTING (VETERINARY USE). NO MORE INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203039 CATGUT PLAIN 1 (5) 75M OTHER SUTURE GAL B. BRAUN SURGICAL, S.A. F1144073 618252

Patients

Seq Age Sex Outcome Treatment
1