FDA Adverse Event Malfunction Summary report: N

ALARIS PC UNIT

MDR report key: 11302835 · Received February 9, 2021

Report

Report Number
2016493-2021-25213
Event Type
Malfunction
Date Received
February 9, 2021
Report Date
December 21, 2020
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Removal / Correction Number
Z-2909-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE WAS EVALUATED AND REPAIRED THROUGH THE SERVICE REPAIR PROCESS. UPON VISUAL INSPECTION THE SERVICE TECHNICIAN NOTED THAT THE ERROR CODE 100.5010 - REFLASHED POLO SOFTWARE. NO POWER TO KEYPAD - REPLACED KEYPAD. A REVIEW OF THE DEVICE HISTORY RECORD SHOWED THE DEVICE HAD A MANUFACTURE DATE OF 08/28/2013. THE REVIEW WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE OF PRODUCT RELEASE FOR DISTRIBUTION. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR (B)(6) WAS PERFORMED WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. BASED ON THE FINDINGS, SERVICE DETERMINED THAT THE PROXIMATE CAUSE OF THE REPORTED ISSUE WAS DUE TO ELECTRICAL FAILURE OF THE ASSY CASE FRONT WITH KEYPAD 8015 LS. A REVIEW OF THE COMPLAINT HISTORY RECORD IN TRACKWISE AND SAP WAS PERFORMED FOR THE (B)(6) WHICH DID NOT CONFIRM SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE FOR THIS CUSTOMER. THERE ARE CAPA #'S NOTED FOR THE FOLLOWING PARTS REPLACED THAT HAVE AN ALREADY EXISTING CAPA. CAPA #: 1120033 FOR PCU UNRESPONSIVE KEYS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE HAD A ALARMED FOR NO APPARENT REASON. NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

THE AFFECTED DEVICE HAS BEEN RECEIVED AND THE INVESTIGATION IS PENDING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE FAILURE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD A ALARMED FOR NO APPARENT REASON. NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
199492 ALARIS PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1