FDA Adverse Event Malfunction Summary report: N

MOME KARDIA

MDR report key: 11296106 · Received February 8, 2021

Report

Report Number
3010112812-2021-00001
Event Type
Malfunction
Date Received
February 8, 2021
Date of Event
January 8, 2021
Report Date
February 4, 2021
Manufacturer
INFOBIONIC INC.
Product Code
DSI
UDI-DI
B690MK10
PMA / PMN Number
K160064
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT AFFECTED BY THE EVENT WAS ENROLLED AND WAS MONITORED IN A CLINICAL RESEARCH STUDY FOR 29 DAYS. TO SUPPRESS BILLING AND USAGE STATISTICS, WHICH WERE NOT DESIRED FOR THIS STUDY, THE DEMO MODE FOR SOFTWARE WAS SET. IN ADDITION TO SUPPRESSING THESE FEATURES STATISTICS, THE DEMO MODE ALSO ALLOWS FOR THE UPLOAD OF SIMULATED PATIENT DATA FOR A PARTICULAR PATIENT DAY, PROVIDED NO OTHER PATIENT DATA HAS BEEN RECEIVED FOR THAT DAY. THE PATIENT TURNED OFF THE DEVICE ON LAST DAY OF THE MONITORING, BEFORE ALL STORED DATA HAD BEEN TRANSMITTED, AND BEGAN THE PROCESS OF RETURNING THE DEVICE TO THE PRACTICE. THE DEVICE HAD NOT YET TRANSMITTED ANY DATA FOR THE LAST DAY WHEN IT WAS TURNED OFF. THREE DAYS LATER, A MONITORING TECHNICIAN NOTICED THAT NO DATA HAD BEEN TRANSMITTED FOR THE LAST DAY. SINCE THERE WAS NO DATA FOR THE LAST DAY YET AVAILABLE, AND THE PRACTICE WAS SET TO DEMO MODE, AN "UPLOAD DATA" BUTTON WAS VISIBLE. THE "UPLOAD DATA" BUTTON ALLOWS FOR SIMULATED/DEMO DATA TO BE UPLOADED FOR ANY PATIENT DAY CONTAINING NO OTHER DATA. A SOFTWARE FEATURE THAT ALLOWS THE TECHNICIAN TO SEE HOW MUCH DATA REMAINS TO BE TRANSMITTED FROM A PATIENT DEVICE WAS NOT USED, AND THE TECHNICIAN PRESSED THE "UPLOAD DATA" BUTTON UNDER THE MISTAKEN ASSUMPTION THAT THIS WOULD CAUSE THE REAL PATIENT DATA TO BE UPLOADED FROM THE PATIENT DEVICE. INSTEAD, SIMULATED DATA WAS UPLOADED FOR THAT DAY, AND AN ATRIAL FIBRILLATION ARRYTHMIA EVENT WAS CREATED AND REPORTED FROM THE SIMULATED DATA. LATER THE DEVICE WAS PHYSICALLY RECEIVED BACK AT THE PRACTICE AND TURNED ON. THE DEVICE THEN BEGAN UPLOADING ITS STORED DATA AND OVERWROTE/REMOVED THE SIMULATED DATA UPLOADED FOR THE LAST DAY. THE MONITORING TECHNICIAN NOTICED THE ERROR WHEN PREPARING THE FINAL REPORT AND NOTIFIED THE PHYSICIAN ON THE RESEARCH TEAM. THE PHYSICIAN THEN COMMUNICATED THE ERROR AND PROVIDED CORRECT DIAGNOSIS TO THE PATIENT AND HIS PHYSICIAN. NO THERAPY WAS ADMINISTERED TO THE PATIENT AS A RESULT OF INITIAL INCORRECT DIAGNOSIS.

Description of Event or Problem · 1

A PRODUCT MALFUNCTION EVENT WAS REPORTED BY HEALTHCARE PROFESSIONAL ON A PHONE CALL REGARDING A PATIENT WHO WAS MONITORED UNDER A CLINICAL RESEARCH STUDY. A SET OF SIMULATED PATIENT DATA (DEMO DATA) HAD BEEN UPLOADED TO ACTUAL PATIENT RECORD DATA INADVERTENTLY AND WAS INTERPRETED AS ACTUAL PATIENT DATA AT THE TIME OF INITIAL DIAGNOSIS. THIS PRODUCT MALFUNCTION CAUSED THE RESEARCH STUDY PATIENT TO BE INCORRECTLY DIAGNOSED WITH ATRIAL FIBRILLATION WHEN INITIAL DIAGNOSIS WAS DONE. THE PATIENT AND HIS PHYSICIAN WERE NOTIFIED OF THIS INCORRECT DIAGNOSIS. ONE OF THE MONITORING TECHNICIANS, WHILE FINALIZING THE PATIENT REPORT LATER, NOTICED THE ERROR AND NOTIFIED THE PHYSICIAN ON THE RESEARCH TEAM. THE PHYSICIAN ON THE RESEARCH TEAM NOTIFIED THE PATIENT AND HIS PHYSICIAN OF THE ERROR AND PROVIDED CORRECT DIAGNOSIS. NO THERAPY WAS ADMINISTERED TO THE PATIENT BASED ON THE INITIAL INCORRECT DIAGNOSIS. NO PATIENT INJURY OR HARM WAS RESULTED AS AN OUTCOME OF THE EVENT. THE SIMULATED DATA WAS LATER SUCCESSFULLY REMOVED AND ORIGINAL CORRECT DATA WAS RESTORED INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193598 MOME KARDIA ARRHYTHMIA DETECTOR AND ALARM DSI INFOBIONIC INC. 01854 092020 B690MK10

Patients

Seq Age Sex Outcome Treatment
1 74 YR