FDA Adverse Event Malfunction Summary report: N

8015 ALARIS SYSTEM PC UNIT

MDR report key: 11283815 · Received February 5, 2021

Report

Report Number
2016493-2021-23329
Event Type
Malfunction
Date Received
February 5, 2021
Report Date
March 3, 2017
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE SOURCE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE PRESENT DATE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). SCAP-UNIT WAS REMEDIATED AND WHEN TESTED BY REMEDIATION FIELD TECH UNIT. EXHIBITED ERROR 133.6080. CHARGE TO RECALL # 911165 PER J HOLBROOK. LOGIC BOARD- DAMAGE. (B)(4). REAR CASE REPLACEMENT NEEDED PER STEVEN BUSTAMANTE, SERVICE TECH. REPAIR WILL BE PERFORMED AT NO ADDITIONAL CHARGE TO THE CUSTOMER UNDER THE PLASTICS REPLACEMENT PROGRAM / RECALL REMEDIATION. (B)(4). FILE REOPENED TO ADD TRACK WISE PR NUMBER TO THE DEVICE DATA FIELD. THIS FILE MAY OR MAY NOT CONTAIN NPI BECAUSE IT WAS INITIALLY CREATED AS A PM / UPGRADE OR RECALL FILE, WHICH DOES NOT REQUIRE THE NPI, AND LATER CONVERTED TO A MAJOR/MINOR REPAIR IN ORDER TO PERFORM THE REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
180898 8015 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1