FDA Adverse Event Injury Summary report: N

PASSY-MUIR TRACHEOSTOMY VENTILATOR SWALLOWING AND SPEAKING VALVE PMV 2001

MDR report key: 11282660 · Received February 4, 2021

Report

Report Number
2024841-2021-00001
Event Type
Injury
Date Received
February 4, 2021
Report Date
February 3, 2021
Manufacturer
PASSY-MUIR, INC.
Product Code
JOH
UDI-DI
10859760005148
PMA / PMN Number
K962714
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT DID NOT INVOLVE MALFUNCTION OF THE PASSY-MUIR VALVE. THE USER FAILED TO FOLLOW MANUFACTURER'S INSTRUCTIONS STATING THAT THE TRACHEOSTOMY CUFF MUST BE COMPLETELY DEFLATED WHEN THE PASSY-MUIR VALVE IS PLACED. THE USER APPEARS TO HAVE APPLIED A MEDICAL DEVICE WITHOUT SUFFICIENT KNOWLEDGE OR UNDERSTANDING OF THE MEDICAL DEVICE. AS THE MANUFACTURER OF THE PASSY-MUIR VALVE, WE TAKE SPECIAL CARE TO PACKAGE EVERY VALVE WITH A COMPREHENSIVE CLINICAL INSTRUCTION BOOKLET, PATIENT HANDBOOK AND CAUTION LABELS. THE CAUTION LABELS ARE BRIGHTLY COLORED AND DESIGNED TO BE PLACED ON THE TRACHEOSTOMY TUBE PILOT LINE WHERE THE CUFF IS INFLATED AND DEFLATED, AS WELL AS AT THE BEDSIDE AND IN THE PATIENT CHART. THIS PRODUCT IS BRIGHTLY COLORED AND MARKED PASSY-MUIR.

Description of Event or Problem · 1

AS REPORTED BY DOCTOR: AT THE END OF A BRIEF PROCEDURE, A CUFFED TRACH TUBE WAS REPLACED IN THE ESTABLISHED TRACHEOTOMY TRACT AND THE PATIENT WAS VENTILATED THROUGH IT. AFTER THE PATIENT WAS BREATHING SPONTANEOUSLY, BUT BEFORE THE ABILITY TO COMMUNICATE EFFECTIVELY RETURNED, THE ASSEMBLY WHICH THE PATIENT HAD WORN INTO THE OR, CONSISTING OF THE (PURPLE) PM VALVE AND A MEANS OF DELIVERING OXYGEN, WAS REPLACED. NOBODY REALIZED THAT EITHER THE VALVE SHOULD HAVE BEEN REMOVED OR THE CUFF SHOULD HAVE BEEN DEFLATED. FAIRLY SOON, THE PATIENT'S RESPIRATORY STATUS DECLINED AND PENEUMOTHORAX WAS IDENTIFIED QUICKLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
178653 PASSY-MUIR TRACHEOSTOMY VENTILATOR SWALLOWING AND SPEAKING VALVE PMV 2001 TRACHEOSTOMY SPEAKING VALVE JOH PASSY-MUIR, INC. PMV 2001 UNKNOWN 10859760005148

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention