ALERE CHOLESTECH LDX ANALYZER (POWER SUPPLY)
Report
- Report Number
- 2027969-2021-00011
- Event Type
- Malfunction
- Date Received
- February 4, 2021
- Report Date
- February 4, 2021
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- JJE
- PMA / PMN Number
- K120615
- Removal / Correction Number
- 2027969-02/14/19-0001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
POWER SUPPLY LOT (1837D) WAS ASSOCIATED WITH FIELD ACTION 2027969-02/14/19-001-C FOR A DEFECTIVE POWER SUPPLY. THE DEFECTIVE POWER SUPPLY RESULTS IN AN INABILITY TO POWER ON THE LDX ANALYZER. NO FURTHER INVESTIGATION WILL BE PURSUED UNDER THIS CASE.
CUSTOMER REPORTS POWER ISSUE ON LDX. POWER CORDS ARE NOT VISIBLY DAMAGED. NO BURNING, NO MELTING, NO SMOKING, NO INJURY. ISSUE WAS RESOLVED WITH AN ALTERNATE LDX POWER CORD. THREE LDX POWER SUPPLIES WERE RECEIVED BY THE INVESTIGATOR ON 1 FEB 2021 AND IT WAS CONFIRMED THE POWER SUPPLIES HAD IDENTIFIER CODE 1837D, WHICH WERE PART OF THE LDX POWER SUPPLY RECALL. NO ADVERSE OUTCOMES REPORTED. THERE WERE THREE LDX ANALYZERS CORRESPONDING WITH THREE AFFECTED LDX POWER SUPPLIES. PLEASE SEE MDRS 2027969-2021-00009 AND 2027969-2021-00010 FOR THE ALTERNATE LDX ANALYZERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 178419 | ALERE CHOLESTECH LDX ANALYZER (POWER SUPPLY) | CHOLESTEROL/LIPIDS ANALYZER | JJE | ALERE SAN DIEGO, INC. | 14-874 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |