FDA Adverse Event Malfunction Summary report: N

TPRLC 133 MP TYPE1 PPS HO

MDR report key: 11280758 · Received February 4, 2021

Report

Report Number
0001825034-2021-00361
Event Type
Malfunction
Date Received
February 4, 2021
Date of Event
January 18, 2021
Report Date
May 12, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWL
UDI-DI
00880304513570
PMA / PMN Number
K110400
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.THE FOLLOWING SECTIONS WERE UPDATED/CORRECTEDUPDATED: D9; G3; H2; H3; H4; H6VISUAL EVALUATION OF THE RETURNED PRODUCT/PHOTOGRAPHS PROVIDED CONFIRMED THERE IS WHITE DEBRIS INSIDE THE STERILE PACKAGING WHICH IS CONSISTENT WITH THE APPEARANCE OF FOAM DEBRIS FROM THE FOAM PACKAGING INSIDE THE STERILE BARRIER. THE REPORTED EVENT IS CONFIRMED.DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND.THE ROOT CAUSE OF THE REPORTED EVENT IS LIKELY TO BE DUE TO TRANSIT DAMAGE CAUSING THE FOAM PACKAGING AND POROUS COATING TO BECOME ABRADED AND SHED.UPON INVESTIGATION, IT HAS BEEN DETERMINED THAT THE DEBRIS IN THE STERILE PACKAGING MEETS THE ACCEPTABLE CRITERIA AND PRODUCT IS CONFORMING TO SPECIFICATIONS. FURTHER, THERE WAS NO ADVERSE EVENT AS A RESULT OF THE REPORTED EVENT. THEREFORE, EVENT IS NO LONGER CONSIDERED REPORTABLE, AND INITIAL REPORT SHOULD BE VOIDED.IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN THE PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED A FOLLOW-UP MDR WILL BE SUBMITTED CONCOMITANT MEDICAL PRODUCTS: ITEM # 51-145170/ TPRLC XR MP T1 PPS/ LOT # 6411825, ITEM# 51-105170/ TPRLC XR T1 PPS/ LOT # 3011823, ITEM # 51-105160/ TPRLC XR T1 PPS /LOT # 6133648, ITEM# 51-107140/ TPRLC 133 MP TYPE1 PPS/ LOT# 6359748, ITEM# 51-145140/ TPRLC XR MP T1 PPS / LOT# 6314883, ITEM# 51-100040/ TPRLC 133 FP TYPE1 PPS / LOT# 6051655. REPORT SOURCE: (B)(6). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS 0001825034-2021-00357, 0001825034-2021-00358, 0001825034-2021-00359, 0001825034-2021-00360, 0001825034-2021-00362, 0001825034-2021-00363.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE WHILE REVIEWING PRODUCT AT THE DISTRIBUTORSHIP, IT WAS NOTED THE PRODUCT WAS FOUND TO HAVE DEBRIS INSIDE THE STERILE PACKAGING. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
179937 TPRLC 133 MP TYPE1 PPS HO PROSTHESIS HIP KWL ZIMMER BIOMET, INC. NI 3743544 00880304513570

Patients

Seq Age Sex Outcome Treatment
1