FDA Adverse Event Malfunction Summary report: N

8100 ALARIS PUMP MODULE

MDR report key: 11278964 · Received February 4, 2021

Report

Report Number
2016493-2021-16347
Event Type
Malfunction
Date Received
February 4, 2021
Report Date
May 22, 2020
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403222054
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). CASE DESCRIPTION: CALLER HAS AN 8100 MODULE GIVING A 242.4030. SN: (B)(4). CASE RESOLUTION: USED KA 12066 ALARIS INFUSION ERROR 242.4030. MOTOR MOVEMENT SEEN. RECOMMENDED TO REPLACE AIL SENSOR. BIOMED WILL CONDUCT REPAIR AND CALL BACK IF FURTHER ASSISTANCE IS NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173951 8100 ALARIS PUMP MODULE PUMP, INFUSION FRN CAREFUSION SD 8100 10885403222054

Patients

Seq Age Sex Outcome Treatment
1