FDA Adverse Event Injury Summary report: N

V60 VENTILATOR

MDR report key: 11276702 · Received February 3, 2021

Report

Report Number
2031642-2021-00426
Event Type
Injury
Date Received
February 3, 2021
Date of Event
January 6, 2021
Report Date
January 21, 2021
Product Code
MNT
PMA / PMN Number
K082660
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

G4: (B)(6) 2021. B4: (B)(6) 2021. A PHILIPS FIELD SERVICE ENGINEER (FSE) EVALUATED THE DEVICE. NO DEVICE SYMPTOM WAS REPORTED DURING THE ADVERSE EVENT, AND NO DEVICE SYMPTOM WAS REPORTED DURING THE DEVICE EVALUATION. SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Additional Manufacturer Narrative · 0

G4: 24FEB2021. B4: 24FEB2021. H11: G5:K102985. H10: PHILIPS FIELD SERVICE ENGINEER (FSE) EVALUATED THE DEVICE AND WAS UNABLE TO DUPLICATE THE SYMPTOM. REVIEW OF THE PROVIDED DIAGNOSTIC REPORT (DRPT) SHOWED NO ERROR CODES WERE GENERATED. NO PARTS WERE REPLACED. THE DEVICE PASSED ALL PERFORMANCE VERIFICATION TESTING AND WAS PLACED BACK INTO USE WITH THE CUSTOMER. THIS REPORTER STATED THAT A PATIENT OF UNKNOWN AGE, GENDER, HEIGHT, AND WEIGHT WAS ADMITTED TO A HOSPITAL ON AN UNKNOWN DATE WITH THE ADMITTING DIAGNOSIS NOT REPORTED. NO RELEVANT MEDICAL HISTORY, RELEVANT PAST DRUG HISTORY OR RELEVANT CONCOMITANT MEDICAL PRODUCTS WERE REPORTED. WHILE ADMITTED ON AN UNKNOWN DATE, THE PATIENT WAS PRESCRIBED VENTILATION THERAPY VIA THE RESPIRONICS V60 VENTILATOR; SOFTWARE VERSION 2.30, PRESCRIPTION, DEVICE SETTINGS, CONFIGURATION, PATIENT CIRCUIT, AND PATIENT INTERFACE NOT REPORTED. WHILE ADMITTED ON (B)(6) 2021, THE PATIENT WAS RECEIVING THERAPY VIA THE V60 DEVICE AND THE PATIENT ¿FLAT LINED¿. NO RELEVANT LABORATORY DATA WAS REPORTED. NO MEDICAL INTERVENTION WAS REPORTED AND THE OUTCOME OF THE PATIENT EXPERIENCING A FLAT LINE IS UNKNOWN. PHILIPS WAS NOT ABLE TO CONFIRM THE REPORTED MALFUNCTION. PHILIPS CONTACTED WITH THE HOSPITAL MANY TIMES, BUT THE HOSPITAL DID NOT PROVIDE ANY DETAILS ABOUT THE INJURY EVENT. NO CONCLUSION CAN BE DRAWN AS THE CUSTOMER DECLINED TO PROVIDE ADDITIONAL INFORMATION. SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Additional Manufacturer Narrative · 1

DATE OF EVENT: (B)(6) 2021. (B)(6) 2021 . SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Description of Event or Problem · 1

A CUSTOMER REPORTED TO PHILIPS THAT WHILE RECEIVING THERAPY FROM THE RESPIRONICS V60 VENTILATOR, THE PATIENT EXPERIENCED A FLAT LINE. THE CUSTOMER REPORTED THAT THE UNIT WAS IN USE ON A PATIENT AT THE TIME OF THE REPORTED ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171358 V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE MNT V60

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER.