FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,8

MDR report key: 11273312 · Received February 3, 2021

Report

Report Number
0002023141-2021-00267
Event Type
Injury
Date Received
February 3, 2021
Date of Event
November 20, 2020
Report Date
May 18, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019263
PMA / PMN Number
K072589
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). TWO IMP,TSV,4.1MM,SBM,8 (TSV4B8) WERE RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED SIGNIFICANT DAMAGE AND FRACTURING AT THE COLLAR. BONE DEBRIS PRESENTED ON THE OUTSIDE THREADS OF THE IMPLANT. ADDITIONALLY, PART OF FRACTURED SCREW STUCK IN THE IMPLANT DRIVE FEATURE. DEVICE HISTORY RECORD (DHR) REVIEWS WERE COMPLETED FOR THE SUBJECT LOT NUMBERS 1230699 AND 1230701. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEWS WERE PERFORMED FOR THE REPORTED LOT NUMBERS (1230699 AND 1230701) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION HAS OCCURRED AND THE REPORTED EVENTS WERE CONFIRMED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). PATIENT IDENTIFIER UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED IMPLANT CAME WITH MOBILITY AND NOTICED THAT THE IMPLANT WAS FRACTURED AND REMOVED IT. ALSO ITS PROSTHIC SCREW WAS FRACTURED. OTHER IMPLANT PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
168349 IMP,TSV,4.1MM,SBM,8 DENTAL IMPLANT DZE ZIMMER DENTAL TSV4B8 1230701 00889024019263

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention