FDA Adverse Event Injury Summary report: N

COCR BALL HEAD 12/14 28 SIZE M 0

MDR report key: 11271432 · Received February 3, 2021

Report

Report Number
3005180920-2021-00086
Event Type
Injury
Date Received
February 3, 2021
Date of Event
January 8, 2021
Report Date
February 3, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805110
PMA / PMN Number
K072857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 20 JANUARY 2021: LOT 100917: 50 ITEMS MANUFACTURED AND RELEASED ON 11-JUNE-2010. EXPIRATION DATE: 2015-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 48 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD. NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED SINCE 2017. ADDITIONAL DEVICE INVOLVED BATCH REVIEW PERFORMED ON 20 JANUARY 2021: LINER: CC E CC LIGHT 01.26.2839HCT FLAT PE HC LINER Ø 28 / C (K103352) LOT 100724: 25 ITEMS MANUFACTURED AND RELEASED ON 06-APRIL-2010. EXPIRATION DATE: 2015-02-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 25 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD. NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED SINCE 2017.

Description of Event or Problem · 1

REVISION SURGERY DUE TO INFECTION 10 YEARS 7 MONTHS AFTER THE PRIMARY AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167548 COCR BALL HEAD 12/14 28 SIZE M 0 HIP METALLIC BALL HEAD JDI MEDACTA INTERNATIONAL SA 01.25.012 100917 07630030805110

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention