FDA Adverse Event Malfunction Summary report: N

SYRINGE 20ML LL BNS

MDR report key: 11270060 · Received February 2, 2021

Report

Report Number
1911916-2021-00082
Event Type
Malfunction
Date Received
February 2, 2021
Date of Event
January 8, 2021
Report Date
January 26, 2021
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(4). INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 301031 AND LOT NUMBER 0003673. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. INVESTIGATION CONCLUSION: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. RATIONALE: THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 20ML LL BNS WAS MISSING SCALE MARKINGS. THIS OCCURRED ON 2212 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 301031, BATCH NO.: 0003673. IT WAS REPORTED THAT SCALE MARKINGS ARE MISSING ON THE SYRINGES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
165028 SYRINGE 20ML LL BNS PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 0003673

Patients

Seq Age Sex Outcome Treatment
1