FDA Adverse Event Malfunction Summary report: N

BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX SYSTEM

MDR report key: 11267821 · Received February 2, 2021

Report

Report Number
1119779-2021-00216
Event Type
Malfunction
Date Received
February 2, 2021
Date of Event
January 7, 2021
Report Date
April 1, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
QJR
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THE COMPLAINT INVESTIGATION FOR INCORRECT LABELLED MASTER MIX POUCHES IN THE BIOGX SARS-COV-2 OSR ASSAY (REF #444213) KIT LOT K20-509 WAS PERFORMED BY BIOGX, PRODUCT MANUFACTURER. THE INVESTIGATION WAS PERFORMED BY THE REVIEW OF THE MANUFACTURING RECORDS, REVIEW OF CUSTOMER¿S PICTURE AND BY THE COMPLAINT¿S HISTORY REVIEW. CUSTOMER REPORTED THAT THAT THE N2 REAGENT POUCH HAD ORANGE FOIL REAGENT AND N1 REAGENT POUCH ALSO HAD ORANGE FOIL REAGENT AND PROVIDED PICTURE FOR INVESTIGATION. THE PICTURES SHOW A POUCH WITH N2 LABEL CONTAINING REAGENT TUBES WITH ORANGE FOIL SEALS. REMINDER THE MASTER MIX TUBES ARE COLOR-CODED: N1 TUBES USE ORANGE FOIL SEALS AND N2 TUBES USE RED FOIL SEALS. BIOGX CONFIRMED THAT MASTER MIX POUCHES EXTERNALLY LABELED AS CONTAINING N2 TUBES WERE CONTAINING N1 TUBES. BIOGX REVIEW OF THE MANUFACTURING RECORDS INDICATED THAT THIS KIT LOT K20-509 WAS AFFECTED BY A HUMAN ERROR. A CHANGE OF PACKAGING SITE HAS CONTRIBUTED TO A HUMAN ERROR DURING THE LABELLING PROCESS OF MASTER MIX POUCHES. BIOGX CONFIRM THE COMPLAINT BASED ON THE REVIEW OF MANUFACTURING RECORD AND THE PICTURE PROVIDED. THERE IS AN INDICATION OF AN INCREASE IN COMPLAINTS FOR INCORRECT LABELLING IN THE BIOGX SARS-COV-2 OSR PRODUCT. THE ROOT CAUSE OF THE CUSTOMER ISSUE WAS A HUMAN ERROR DURING LABELLING PROCESS. BIOGX PRODUCT MANUFACTURER HAS INITIATED A PREVENTIVE AND CORRECTIVE ACTION PLAN. BD HAS ALSO INITIATED A CAPA (2208364).

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX¿ SYSTEM THE N2 REAGENT IN POUCH HAD ORANGE CAP INSTEAD OF RED. (B)(4).

Additional Manufacturer Narrative · 1

EUA# (B)(4). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE WITH BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX¿ SYSTEM THE N2 REAGENT IN POUCH HAD ORANGE CAP INSTEAD OF RED. EUA# (B)(4).