INTELLIVUE MX40 802.11A/B/G
Report
- Report Number
- 1218950-2021-00689
- Event Type
- Malfunction
- Date Received
- February 2, 2021
- Report Date
- January 27, 2021
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- DSI
- PMA / PMN Number
- K113125
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
H10: THE ACTUAL DEVICE WAS RETURNED TO PHILIPS BENCH WHERE THE TECHNICIAN WAS UNABLE TO REPLICATE THE CUSTOMER'S ALLEGATION OF NO SOUND. THE TECHNICIAN CONFIRMED THE DEVICE HAD SPEAKER MALFUNCTION ALARM AND LOG MESSAGES; HOWEVER, THE AUDIO WAS WORKING. THERE WAS CORROSION ON THE SYSTEM BOARD NEAR THE SPEAKER CONNECTOR, WHICH CAUSED THE SPEAKER ALARM MESSAGES. CORROSION IS KNOWN TO BE CAUSED BY IMPROPER CLEANING/DISINFECTION OF THE MX40/MX4J PMW AND/OR USE OF UNAPPROVED CLEANING AND DISINFECTING AGENTS. SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.
THE CUSTOMER REPORTED A SPEAKER MALFUNCTION ERROR MESSAGE AND A LOSS OF AUDIO ON THE MX40 TELEMETRY DEVICE. IT WAS REPORTED THE DEVICE WAS NOT IN CLINICAL USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 167417 | INTELLIVUE MX40 802.11A/B/G | PATIENT MONITOR | DSI | PHILIPS MEDICAL SYSTEMS | 865352 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |