CD HORIZON SPINAL SYSTEM
Report
- Report Number
- 1030489-2021-00117
- Event Type
- Malfunction
- Date Received
- February 2, 2021
- Report Date
- February 2, 2021
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- KWP
- PMA / PMN Number
- SEE H10
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT IS NOT MARKETED IN US. 510K FOR SIMILAR PRODUCT MARKETED IN US WITH CATALOGUE# 70268580 IS K050439. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS OF SCREW BACK OUT AND ROD BROKEN. IT WAS A REVISION SURGERY. INITIAL SURGERY IS DONE ON 2016, TLIF WAS PERFORMED ON ONE INTERVERTEBRAL DISC. ON (B)(6) 2017, REOPERATION WAS PERFORMED ON T4-S2AI. ON (B)(6) 2017, REPLACEMENT OF VERTEBRAL BODY WAS PERFORMED BECAUSE L4 WAS COLLAPSED. ON (B)(6) 2019, S1 AND S2AI SCREWS WERE REPLACED (SET SCREW BACKED OUT). ON (B)(6) 2019, S1 AND S2AI SCREW REPLACEMENT + EXTENSION TO IL WAS DONE BECAUSE SET SCREW BACKED OUT. IT WAS REPORTED THAT THE SET SCREW ON THE C SIDE OF THE MRC O / C INSERTED BETWEEN THE LEFT L2 AND L3 BACKED OUT. ROD WAS BROKEN BETWEEN LEFT L3-4.LEGACY CLOSED MAS F8.5X80 SCREW OF THE LEFT S2AI WAS BROKEN ON THE BASE. ROD WAS BROKEN BETWEEN THE RIGHT L2-3.SET SCREW OF THE RIGHT S1 BACKED OUT. LEGACY CLOSED MAS F8.5X80 SCREW OF THE RIGHT S2AI WAS BROKEN ON THE BASE. HENCE ON (B)(6) 2021, ALL THE RODS, CONNECTORS, AND LEGACY CLOSED MAS WERE REMOVED AND REPLACED, AND THE OPERATION WAS COMPLETED. THERE WERE FRAGMENTS OF THE IMPLANTS REMAINED IN THE PATIENT BODY. TWO S2AI LEGACY CLOSED MAS8.5X80 SCREWS REMAINING IN THE SACRAL ILIAC PART.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163662 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | WARSAW ORTHOPEDICS | GX70468580 | 0272863W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention |