8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2021-22032
- Event Type
- Malfunction
- Date Received
- February 1, 2021
- Report Date
- March 1, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
CASE #01040226 - 8015 DID NOT COMMUNICATE WITH SYSTEM MAINTENANCE SOFTWARE THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT. (B)(4). CASE DESCRIPTION: (B)(6) HAD A PCU8015 DID NOT COMMUNICATE WITH SYSTEM MAINTENANCE SOFTWARE. PCU SN (B)(4). FAILURE DEVICE TYPE: FAILURE PROBLEM TYPE: FAILURE MODE: CASE RESOLUTION: I REVIEWED THE COMMUNICATION CABLE SET UP WITH (B)(6) AND RECOMMENDED HIM TO REPLACE SIO BOARD ASSY. ASSISTED WITH A PART NUMBER. (B)(6) WILL REPLACE SIO BOARD ASSY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159815 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |