FDA Adverse Event Malfunction Summary report: N

BD CALIBRITE 3 BEADS

MDR report key: 11259331 · Received February 1, 2021

Report

Report Number
2916837-2021-00040
Event Type
Malfunction
Date Received
February 1, 2021
Date of Event
January 12, 2021
Report Date
February 16, 2021
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
GKZ
UDI-DI
00382903404865
PMA / PMN Number
K973483
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION OF THE COMPLAINT, IT HAS BEEN DETERMINED THAT THE PREVIOUSLY SUBMITTED MFR REPORT# 2916837-2021-00040 WAS SENT IN ERROR. THERE WERE NO PATIENT SAMPLES INVOLVED, AND REAGENT LOT-TO-LOT VARIATION IS NOT CONSIDERED TO BE A REPORTABLE MALFUNCTION.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD CALIBRITE¿ 3 BEADS THERE WAS A POOR FL3 PERFORMANCE. CONFIRMATORY TESTING PERFORMED RUNNING TEST SIMULTANEOUSLY ON TO RULE OUT MACHINE ISSUE. NEW LOT OF BEADS PASSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NATURE OF INCIDENT: POOR FL3 PERFORMANCE. THE AFFECTED CALIBRITE BEAD LOT NUMBER IS 0002577. TESTED SAME CALIBRITE BEAD ON 2 UNIT FACSCALIBUR SIMULTANEOUSLY (MACHINE A & MACHINE B) TO RULE OUT MACHINE PROBLEM. RUN FACSCOMP USING DIFFERENT LOT OF CALIBRITE BEAD #0212458 AND IT SHOW PASS.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED WHILE TESTING WITH BD CALIBRITE¿ 3 BEADS THERE WAS A POOR FL3 PERFORMANCE. CONFIRMATORY TESTING PERFORMED RUNNING TEST SIMULTANEOUSLY ON TO RULE OUT MACHINE ISSUE. NEW LOT OF BEADS PASSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NATURE OF INCIDENT: POOR FL3 PERFORMANCE. THE AFFECTED CALIBRITE BEAD LOT NUMBER IS 0002577 TESTED SAME CALIBRITE BEAD ON 2 UNIT FACSCALIBUR SIMULTANEOUSLY (MACHINE A & MACHINE B) TO RULE OUT MACHINE PROBLEM. RUN FACSCOMP USING DIFFERENT LOT OF CALIBRITE BEAD #0212458 AND IT SHOW PASS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154604 BD CALIBRITE 3 BEADS COUNTER, DIFFERENTIAL CELL GKZ BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES 340486 0002577 00382903404865

Patients

Seq Age Sex Outcome Treatment
1