BD CALIBRITE 3 BEADS
Report
- Report Number
- 2916837-2021-00040
- Event Type
- Malfunction
- Date Received
- February 1, 2021
- Date of Event
- January 12, 2021
- Report Date
- February 16, 2021
- Manufacturer
- BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
- Product Code
- GKZ
- UDI-DI
- 00382903404865
- PMA / PMN Number
- K973483
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AFTER FURTHER EVALUATION OF THE COMPLAINT, IT HAS BEEN DETERMINED THAT THE PREVIOUSLY SUBMITTED MFR REPORT# 2916837-2021-00040 WAS SENT IN ERROR. THERE WERE NO PATIENT SAMPLES INVOLVED, AND REAGENT LOT-TO-LOT VARIATION IS NOT CONSIDERED TO BE A REPORTABLE MALFUNCTION.
IT WAS REPORTED WHILE TESTING WITH BD CALIBRITE¿ 3 BEADS THERE WAS A POOR FL3 PERFORMANCE. CONFIRMATORY TESTING PERFORMED RUNNING TEST SIMULTANEOUSLY ON TO RULE OUT MACHINE ISSUE. NEW LOT OF BEADS PASSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NATURE OF INCIDENT: POOR FL3 PERFORMANCE. THE AFFECTED CALIBRITE BEAD LOT NUMBER IS 0002577. TESTED SAME CALIBRITE BEAD ON 2 UNIT FACSCALIBUR SIMULTANEOUSLY (MACHINE A & MACHINE B) TO RULE OUT MACHINE PROBLEM. RUN FACSCOMP USING DIFFERENT LOT OF CALIBRITE BEAD #0212458 AND IT SHOW PASS.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED WHILE TESTING WITH BD CALIBRITE¿ 3 BEADS THERE WAS A POOR FL3 PERFORMANCE. CONFIRMATORY TESTING PERFORMED RUNNING TEST SIMULTANEOUSLY ON TO RULE OUT MACHINE ISSUE. NEW LOT OF BEADS PASSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NATURE OF INCIDENT: POOR FL3 PERFORMANCE. THE AFFECTED CALIBRITE BEAD LOT NUMBER IS 0002577 TESTED SAME CALIBRITE BEAD ON 2 UNIT FACSCALIBUR SIMULTANEOUSLY (MACHINE A & MACHINE B) TO RULE OUT MACHINE PROBLEM. RUN FACSCOMP USING DIFFERENT LOT OF CALIBRITE BEAD #0212458 AND IT SHOW PASS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154604 | BD CALIBRITE 3 BEADS | COUNTER, DIFFERENTIAL CELL | GKZ | BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES | 340486 | 0002577 | 00382903404865 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |