FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 11259278 · Received February 1, 2021

Report

Report Number
2024168-2021-00753
Event Type
Injury
Date Received
February 1, 2021
Date of Event
January 17, 2021
Report Date
April 28, 2021
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648013089
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

VISUAL AND FUNCTIONAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED DEPLOYMENT ISSUE COULD NOT BE TESTED DUE TO THE CONDITION OF THE RETURNED DEVICE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY IDENTIFIED NO SIMILAR INCIDENTS FROM THIS LOT. THE INVESTIGATION WAS UNABLE TO DETERMINE A CONCLUSIVE CAUSE FOR THE REPORTED DIFFICULTIES; HOWEVER, THE TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. B3: DATE OF PROCEUDRE D4; THE LOT #0101441 CHANGED TO LOT#0090641. H6: TYPE OF INVESTIGATION CODE 4115 REMOVED, INVESTIGATION FINDINGS CODE 3221 REMOVED D10, H3 - IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD NOT BE RETURNING FOR ANALYSIS. SUBSEQUENT INFORMATION REVEALED THAT THE DEVICE WAS RETURNED FOR EVALUATION.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY IDENTIFIED NO SIMILAR INCIDENTS FROM THIS LOT. THE INVESTIGATION WAS UNABLE TO DETERMINE A CONCLUSIVE CAUSE FOR THE REPORTED DIFFICULTIES; HOWEVER, THE TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.

Description of Event or Problem · 1

IT WAS REPORTED THAT A VESSEL PUNCTURE CLOSURE WITH A PROGLIDE DEVICE WAS ATTEMPTED IN AN UNSPECIFIED VESSEL RELATIVE TO AN ABDOMINAL AORTIC ANEURYSM PROCEDURE. REPORTEDLY, THE PROGLIDE DEVICE FAILED TO DEPLOY PROPERLY. THE METHOD USED TO ACHIEVE HEMOSTASIS WAS NOT REPORTED. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154587 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 12673-03 0090641 08717648013089

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention