FDA Adverse Event Injury Summary report: N

MLRY-HD LAT POR FMRL 12X165MM

MDR report key: 11246044 · Received January 28, 2021

Report

Report Number
0001825034-2021-00283
Event Type
Injury
Date Received
January 28, 2021
Date of Event
November 9, 2015
Report Date
January 28, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
K030055
Removal / Correction Number
NI
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). REPORTED EVENT WAS CONFIRMED VIA OPERATIVE NOTES REVIEWED BY A HEALTH CARE PROFESSIONAL. SEM ANALYSIS IDENTIFIED: THE TAPER OF THE RETURNED DEVICE WAS REVIEWED VIA OPTICAL MICROSCOPY RESULTING IN A CONSENSUS MODIFIED GOLDBERG SCORE OF 4. A SCORE OF 4 CORRESPONDS TO, DAMAGE OVER THE MAJORITY (>50%) OF THE SURFACE WITH SEVERE CORROSION ATTACK AND ABUNDANT CORROSION DEBRIS. REVIEW OF COMPLAINT HISTORY FOR ITEM# 11-104212 LOT# 213430 IDENTIFIED ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED ITEM(S) AND PART AND LOT COMBINATION(S). COMPLAINTS ARE MONITORED THROUGH MONTHLY COMPLAINT REVIEW (REFERENCE (B)(4) AND (B)(4)) IN ORDER TO IDENTIFY POTENTIAL ADVERSE TRENDS. ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT RIGHT TOTAL HIP REVISION PROCEDURE APPROXIMATELY ELEVEN YEARS¿ POST-IMPLANTATION DUE TO PAIN AND METALLOSIS WAS NOTED DURING THE PROCEDURE. THE MEDICAL RECORDS STATE: THE COMPONENT WAS SLIGHTLY RETROVERTED WITH A PROMINENT ANTERIOR FLANGE WHICH MAY HAVE CAUSED SOFT TISSUE IRRITATION AND MAY HAVE BEEN THE SOURCE OF GROIN PAIN. THERE WERE NO COMPLICATIONS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142858 MLRY-HD LAT POR FMRL 12X165MM PROSTHESIS, HIP LPH ZIMMER BIOMET, INC. N/A 213430

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R