FDA Adverse Event Injury Summary report: N

NA

MDR report key: 11245222 · Received January 28, 2021

Report

Report Number
2015691-2021-01038
Event Type
Injury
Date Received
January 28, 2021
Date of Event
January 4, 2021
Report Date
February 23, 2021
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H10. ADDITIONAL MANUFACTURER NARRATIVE: UPDATED SECTIONS D4 EXPIRATION DATE, H4, AND H6. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWS THAT THIS DEVICE MET ALL MANUFACTURING SPECIFICATIONS FOR PRODUCT RELEASE PRIOR TO DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT.

Additional Manufacturer Narrative · 1

PMA/510(K): 8300ACA- THIS DEVICE IS NOT SOLD OR MARKETED IN THE U.S.; HOWEVER, IT IS SIMILAR TO DEVICE MODEL 8300AB, EDWARDS INTUITY ELITE VALVE (PMA P150036). THE DEVICE WAS NOT RETURNED TO EDWARDS FOR EVALUATION. CALCIFICATION IS A WELL-RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. MANY FACTORS CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION. THESE INCLUDE PATIENT FACTORS (AGE, DISEASE STATE, PHARMACOLOGICAL INTERVENTION, ETC.), AND MECHANICAL STRESS RELATED TO THE VALVE'S HEMODYNAMIC PERFORMANCE. THOUGH NUMEROUS STUDIES HAVE BEEN CONDUCTED ON PREVENTIVE CALCIFICATION STRATEGIES IN BIOPROSTHETIC HEART VALVES, THE CAUSES OF CALCIFICATION ARE NOT FULLY UNDERSTOOD AND THERE ARE STILL NO MECHANISMS OR MEDICAL THERAPIES WHICH FULLY PREVENT BIOPROSTHESES FROM CALCIFYING. ALTHOUGH PATIENT FACTORS ARE BELIEVED TO PLAY A CRUCIAL ROLE IN THE DEVELOPMENT IN BIOPROSTHETIC TISSUE CALCIFICATION, THE UNDERLINE MECHANISM IS STILL NOT FULLY UNDERSTOOD. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED; HOWEVER, IT IS LIKELY THAT PATIENT RELATED FACTORS AND THE PROGRESSION OF THE UNDERLYING VALVULAR DISEASE PATHOLOGY CONTRIBUTED TO THE EVENT. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MDR WILL BE SUBMITTED. EDWARDS LIFESCIENCES WILL CONTINUE TO MONITOR ALL REPORTED EVENTS. NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME.

Description of Event or Problem · 1

(B)(4): IT WAS REPORTED THAT A CLINICAL TRIAL PATIENT WITH A 19MM 8300ACA AORTIC VALVE IMPLANTED FOR SEVEN YEARS, NINE MONTHS UNDERWENT REDO AVR DUE TO CALCIFICATION, SEVERE STENOSIS, AND DECREASED LEAFLET EXCURSION DETECTED BY ECHO. A 21MM 8300AB VALVE WAS IMPLANTED IN REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
140222 NA REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 8300ACA NA

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| L| R