FDA Adverse Event Malfunction Summary report: N

ELEVATE SPINAL SYSTEM

MDR report key: 11242706 · Received January 28, 2021

Report

Report Number
1030489-2021-00105
Event Type
Malfunction
Date Received
January 28, 2021
Date of Event
January 7, 2021
Report Date
June 15, 2021
Manufacturer
WARSAW ORTHOPEDICS
Product Code
MAX
UDI-DI
00643169430433
PMA / PMN Number
K172199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

PRODUCT ANALYSIS OF PRODUCT#7770723 ; LOT# 0758071W ANALYSIS SUMMARY: VISUAL AND OPTICAL EXAMINATION REVIEW CONFIRMS IMPLANT THREADS DAMAGED. WITNESS MARKS AND MATERIAL DEFORMATION NOTED ON THE THRU HOLE OF THE PEEK ANTERIOR RAMP. THESE OBSERVATIONS ARE CONSISTENT WITH OVERLOAD OF THE IMPLANT. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING INTRA-OP FOR A PATIENT WITH UNKNOWN SPINAL THERAPY. IT WAS REPORTED THAT, THE CAGE WOULD NOT EXPAND OR CONTRACT. IT WAS MENTIONED THAT IT WAS REMOVED AND ANOTHER ONE OPENED AND IMPLANTED. IT WAS MENTIONED AS UPON PREPARING THE IMPLANT ON THE INSERTER, THE EXPANDING SECTION WOULD NOT LOCK TIGHT IN THE CLOSED POSITION. ONCE INSERTED THE DEVICE DID NOT OPENED, DOCTOR REMOVED IT AND OPENED A NEW IMPLANT. THERE WERE NO SYMPTOMS TO PATIENT REPORTED AS A RESULT OF THIS EVENT. THERE WAS NO ADDITIONAL SURGERY OR TREATMENT PERFORMED AS A RESULT OF THIS EVENT. THERE WERE NO FURTHER COMPLICATIONS TO PATIENT OR PHYSICIAN WERE REPORTED OR ANTICIPATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139438 ELEVATE SPINAL SYSTEM INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR MAX WARSAW ORTHOPEDICS 7770723 0758071W 00643169430433

Patients

Seq Age Sex Outcome Treatment
1 48 YR